MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1259/bjr.20220373

Медицинская библиотека · The British journal of radiology · 2023

Prediction of dose deposition matrix using voxel features driven machine learning approach.

Jiao Shengxiu, Zhao Xiaoqian, Yao Shuzhan

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

A dose deposition matrix (DDM) prediction method using several voxel features and a machine learning (ML) approach is proposed for plan optimization in radiation therapy.

METHODS

Head and lung cases with the inhomogeneous medium are used as training and testing data. The prediction model is a cascade forward backprop neural network where the input is the features of the voxel, including 1) voxel to body surface distance along the beamlet axis, 2) voxel to beamlet axis distance, 3) voxel density, 4) heterogeneity corrected voxel to body surface distance, 5) heterogeneity corrected voxel to beamlet axis, and (6) the dose of voxel obtained from the pencil beam (PB) algorithm. The output is the predicted voxel dose corresponding to a beamlet. The predicted DDM was used for plan optimization (ML method) and compared with the dose of MC-based plan optimization (MC method) and the dose of pencil beam-based plan optimization (PB method). The mean absolute error (MAE) value was calculated for full volume relative to the dose of the MC method to evaluate the overall dose performance of the final plan.

RESULTS

For patient with head tumor, the ML method achieves MAE value 0.49 × 10-4 and PB has MAE 1.86 × 10-4. For patient with lung tumor, the ML method has MAE 1.42 × 10-4 and PB has MAE 3.72 × 10-4. The maximum percentage difference in PTV dose coverage (D98) between ML and MC methods is no more than 1.2% for patient with head tumor, while the difference is larger than 10% using the PB method. For patient with lung tumor, the maximum percentage difference in PTV dose coverage (D98) between ML and MC methods is no more than 2.1%, while the difference is larger than 16% using the PB method.

CONCLUSIONS

In this work, a reliable DDM prediction method is established for plan optimization by applying several voxel features and the ML approach. The results show that the ML method based on voxel features can obtain plans comparable to the MC method and is better than the PB method in achieving accurate dose to the patient, which is helpful for rapid plan optimization and accurate dose calculation.

ADVANCES IN KNOWLEDGE: Establishment of a new machine learning method based on the relationship between the voxel and beamlet features for dose deposition matrix prediction in radiation therapy.

Для цитирования: Jiao Shengxiu, Zhao Xiaoqian, Yao Shuzhan. Prediction of dose deposition matrix using voxel features driven machine learning approach.. The British journal of radiology. 2023. DOI: 10.1259/bjr.20220373.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.