Laryngeal damage due to an unexpectedly large and inappropriately designed cuffed pediatric tracheal tube in a 13-month-old child.

Dillier Claudia M, Trachsel Daniel, Baulig Werner, Gysin Claudine, Gerber Andreas C, Weiss Markus

Клинический случайСтатьяПеревод в очереди

Аннотация

PURPOSE: To present a case of laryngeal damage in an infant caused by a too large and inappropriately designed cuffed tracheal tube.

CLINICAL FEATURES: A 13-month-old child undergoing cardiac surgery was intubated with an uncuffed endotracheal tube with an internal diameter (ID) of 4.0 mm. Because of an important air leak around the tracheal tube during mechanical ventilation, a cuffed endotracheal tube ID 4.0 mm was inserted. The air leak with the tube cuff not inflated was acceptable at 25 cm H2O airway pressure. After extubation on the third postoperative day, the patient showed increasing stridor and respiratory deterioration. Fibreoptic laryngoscopy of the spontaneously breathing patient showed a large intra-laryngeal web. After surgical removal of the web, the child rapidly recovered and was discharged from the hospital on the 12th postoperative day. Inspection of the 4.0 mm (ID) cuffed tracheal tube revealed a cuff positioned inappropriately high and an increase of 0.7 mm in outer tube diameter compared to the 4.0 mm (ID) uncuffed tracheal tube from the same manufacturer. The tube cuff is likely to be situated within the larynx when placed in accordance to insertion depth formulas or radiological criteria, as used for uncuffed tracheal tubes in children.

CONCLUSION

The larger than expected tracheal tube with its intra-laryngeal cuff position in a 13-month-old child likely caused mucosal damage and an inflammatory reaction within the larynx resulting in granulation tissue formation and fibrous healing around the tracheal tube.

Для цитирования: Dillier Claudia M, Trachsel Daniel, Baulig Werner, Gysin Claudine, Gerber Andreas C, Weiss Markus. Laryngeal damage due to an unexpectedly large and inappropriately designed cuffed pediatric tracheal tube in a 13-month-old child.. Canadian journal of anaesthesia = Journal canadien d'anesthesie. 2004. DOI: 10.1007/bf03018551.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.