Nasopharynx is well-suited for core temperature measurement during hypothermia therapy.

Hine Kotaro, Hosono Shigeharu, Kawabata Ken, Miyabayashi Hiroshi, Kanno Keichi, Shimizu Masaki, Takahashi Shori

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Rectal temperature is commonly used as the core temperature during therapeutic hypothermia therapy in neonates with hypoxic-ischemic encephalopathy (HIE). The purpose of this study was to examine whether nasopharyngeal temperature could serve as a substitute for rectal temperature.

METHODS

We prospectively investigated 40 neonates with HIE who underwent therapeutic hypothermia by selective head cooling, which involved cooling the body to 34°C for 72 h. During this period, nasopharyngeal temperature was measured and compared with rectal temperature every hour.

RESULTS

For 40 neonates included in this study, the mean rectal and nasopharyngeal temperatures were 34.3 ± 0.4°C (n = 2920) and 34.3 ± 0.4°C (n = 2920), respectively. Nasopharyngeal temperature strongly correlated with rectal temperature (R2 = 0.623, P < 0.0001) and magnitude of the mean difference between nasopharyngeal and rectal temperature varied little during the 72 h of therapeutic hypothermia.

CONCLUSIONS

Nasopharyngeal temperature in neonates with perinatal HIE undergoing therapeutic hypothermia may be a suitable substitute for rectal temperature.

Для цитирования: Hine Kotaro, Hosono Shigeharu, Kawabata Ken, Miyabayashi Hiroshi, Kanno Keichi, Shimizu Masaki, Takahashi Shori. Nasopharynx is well-suited for core temperature measurement during hypothermia therapy.. Pediatrics international : official journal of the Japan Pediatric Society. 2016. DOI: 10.1111/ped.13046.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.