Early Experience With a Novel Dissection-Specific Stent-Graft to Prevent Distal Stent-Graft-Induced New Entry Tears After Thoracic Endovascular Repair of Chronic Type B Aortic Dissections.

Burdess Anne, D'Oria Mario, Mani Kevin, Tegler Gustaf, Lindström David, Mogensen John, Kölbel Tilo, Wanhainen Anders

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The aim was to report short and mid-term outcomes of a novel, investigational, dissection-specific stent-graft (DSSG), specifically designed to address the features of chronic type B aortic dissection (CTBAD) and reduce the risk of distal stent-graft-induced new entry tears (dSINE).

MATERIALS AND

METHODS

A retrospective single center cohort study of all patients undergoing TEVAR with the DSSG for CTBAD from January 1, 2017 to January 31, 2020. The DSSG, which is a modified stent-graft based on the Cook Zenith Alpha Thoracic platform, has no proximal barbs, and a customized longer body length with substantial taper. The second and third distal Z-stents are sited internally to avoid any contact of the metal skeleton with the dissection membrane and have reduced radial force, while the most distal stent was removed creating a distal 30 mm unsupported Dacron graft.

RESULTS

Sixteen patients (13 males, 3 females) with a median age of 66 years (range 31-79 years) underwent elective TEVAR of CTBAD using the DSSG. Six patients (38%) had an underlying connective tissue disorder. The median tapering was 10 mm (range 4 mm-21 mm) and median length 270 mm (range 210-380 mm). Technical success was achieved in all but one case (96%). One patient died within 30 days, due to retrograde type A dissection with cardiac tamponade. The 30-day rate of stroke, spinal cord ischemia, and re-interventions was 0%. After median imaging follow-up time of 17 months (range 1-31 months), one patient developed a dSINE 4 months after the index procedure. After median survival follow-up of 23 months (range 2-35 months), one late death occurred due to traumatic brain injury, while no aortic-related death occurred during follow-up. Complete false lumen (FL) thrombosis was achieved in 9 patients while the remaining 6 showed partial FL thrombosis. No instances of diameter increase at the level oftreated aortic segment were noted with serial measurements showing either stable (n = 7) or decreased (n = 8) maximal transverse diameter.

CONCLUSIONS

Use of a novel DSSG with low radial force for TEVAR in the setting of CTBAD is safe and feasible. This early real-world experience shows promising mid-term effectiveness with low rates of dSINE or unplanned re-interventions and satisfactory aortic remodeling during follow-up. Longer follow-up is needed, however, before any firm conclusions can be drawn.

Для цитирования: Burdess Anne, D'Oria Mario, Mani Kevin, Tegler Gustaf, Lindström David, Mogensen John, Kölbel Tilo, Wanhainen Anders. Early Experience With a Novel Dissection-Specific Stent-Graft to Prevent Distal Stent-Graft-Induced New Entry Tears After Thoracic Endovascular Repair of Chronic Type B Aortic Dissections.. Annals of vascular surgery. 2021. DOI: 10.1016/j.avsg.2021.10.048.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.