MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/jmf.11.2.80.83

Медицинская библиотека · The journal of maternal-fetal & neonatal medicine : the official journal of the European Association of Perinatal Medicine, the Federation of Asia and Oceania Perinatal Societies, the International Society of Perinatal Obstetricians · 2002

Is the correlation between fetal oxygen saturation and blood pH sufficient for the use of fetal pulse oximetry?

Rijnders R J P, Mol B W J, Reuwer P J H M, Drogtrop A P, Vernooij M M A, Visser G H A

Клинический случайСтатьяПеревод в очереди

Аннотация

OBJECTIVES

Fetal pulse oximetry was performed during labor in high-risk cases for fetal distress to determine the diagnostic value of this method.

METHODS

The fetal SpO2 values were blinded from the obstetrician so that these values did not influence clinical decisions. Mean and lowest SpO2 measurements for the last 30 min prior to either fetal scalp blood sampling or delivery were correlated with scalp pH or pH from the umbilical artery.

RESULTS

No significant correlation was found between pH and mean fetal oxygen saturation (correlation coefficient -0.02, p = 0.9). There was no significant correlation between pH and lowest fetal oxygen saturation (correlation coefficient 0.04, p = 0.84). Concerning the feasibility of the method, we found that only 23 of 65 included patients were suitable for analysis; in 20% of cases, we were not able to perform a SpO2 measurement.

CONCLUSIONS

None of three cases with pH below 7.05 would have been detected using mean SpO2 over the last 30 min prior to fetal scalp blood sampling or delivery. Only one case would have been detected using the lowest SpO2 measurement over this period. We conclude that fetal SpO2 measurements during labor are of poor diagnostic value, with a disappointing feasibility and therefore are not ready for implementing into daily clinical practice.

Для цитирования: Rijnders R J P, Mol B W J, Reuwer P J H M, Drogtrop A P, Vernooij M M A, Visser G H A. Is the correlation between fetal oxygen saturation and blood pH sufficient for the use of fetal pulse oximetry?. The journal of maternal-fetal & neonatal medicine : the official journal of the European Association of Perinatal Medicine, the Federation of Asia and Oceania Perinatal Societies, the International Society of Perinatal Obstetricians. 2002. DOI: 10.1080/jmf.11.2.80.83.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.