MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10840-025-02021-y

Медицинская библиотека · Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2025

Redo ablation procedures to treat recurrent atrial arrhythmias via a pentaspline pulsed field ablation catheter: a prospective, multicenter experience.

Cespón-Fernández María, Della Rocca Domenico G, Magnocavallo Michele, Betancur Andrés, Lombardo Ilenia, Pannone Luigi, Vetta Giampaolo, Sorgente Antonio, Polselli Marco, Audiat Charles, Del Monte Alvise, Combes Stéphane, Marcon Lorenzo, Overeinder Ingrid, Nakasone Kazutaka, Mouram Sahar, Mohanty Sanghamitra, Bianchi Stefano, Almorad Alexandre, Sieira Juan, Bala Gezim, Ströker Erwin, Rossi Pietro, Sarkozy Andrea, Boveda Serge, Natale Andrea, de Asmundis Carlo, Chierchia Gian-Battista, HRMC Investigators

Оценочное исследованиеСтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND

Patients undergoing atrial fibrillation (AF) catheter ablation may require redo procedures involving pulmonary vein (PV) re-isolation and/or ablation of extra-PV sites. Pulsed field ablation (PFA) offers a highly selective energy source for cardiac tissue, with the potential to reduce collateral damage to adjacent structures. This study aimed to evaluate the feasibility and efficacy of redo ablation using a pentaspline PFA system.

METHODS

Patients undergoing redo procedures with a pentaspline PFA system at three international centers were enrolled. A workflow was established based on rhythm at presentation: sinus rhythm (Group 1), atrial flutter/atrial tachycardia (Group 2), or AF (Group 3). Propensity score matching was used for comparison between PFA- and RF-based redo ablations.

RESULTS

A total of 117 patients were included (Group 1: 64, Group 2: 18, Group 3: 35). PV re-isolation was required in 71.9% and 72.2% of Group 1 and 2 patients, respectively. PFA terminated all cases of non-cavotricuspid isthmus dependent flutter and 45.7% of cases of AF. One major complication (0.9%; frontal cerebral hematoma) was documented. Freedom from atrial tachyarrhythmias at 12 months was 78.3% (95% CI 69.6-84.8%) without statistically significant differences among groups (Group 1: 85.7%; Group 2: 77%; Group 3: 65.5%; p = 0.053). PFA led to similar arrhythmia freedom compared to RF, but with significantly shorter procedural and dwelling times.

CONCLUSION

The adoption of a pentaspline PFA system for repeat ablation procedures was feasible, safe, and effective at 1-year follow-up. No clinical differences were observed between PFA and RF; however, redo PFA cases were significantly shorter.

Для цитирования: Cespón-Fernández María, Della Rocca Domenico G, Magnocavallo Michele, Betancur Andrés, Lombardo Ilenia, Pannone Luigi, Vetta Giampaolo, Sorgente Antonio, Polselli Marco, Audiat Charles, Del Monte Alvise, Combes Stéphane, Marcon Lorenzo, Overeinder Ingrid, Nakasone Kazutaka, Mouram Sahar, Mohanty Sanghamitra, Bianchi Stefano, Almorad Alexandre, Sieira Juan, Bala Gezim, Ströker Erwin, Rossi Pietro, Sarkozy Andrea, Boveda Serge, Natale Andrea, de Asmundis Carlo, Chierchia Gian-Battista, HRMC Investigators. Redo ablation procedures to treat recurrent atrial arrhythmias via a pentaspline pulsed field ablation catheter: a prospective, multicenter experience.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2025. DOI: 10.1007/s10840-025-02021-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.