Development and Validation of a Free Carbamazepine Assay on an Automated Chemistry Analyzer.
Аннотация
BACKGROUND
Carbamazepine is an effective drug for treating seizures and trigeminal neuralgia. Therapeutic drug monitoring of free carbamazepine in serum can be useful in situations that drug--protein binding is altered to guide regimen adjustment and to aid in the diagnosis of clinical toxicity.
METHODS
Separation of the nonprotein bound carbamazepine was achieved via ultrafiltration through a molecular weight cut-off filter. A method for free carbamazepine measurement was developed on the automated cobas chemistry analyzers (Roche Diagnostics) by modifying the Carbamazepine Gen 4 assay (Roche Diagnostics). Assay performance characteristics were established including precision, accuracy, reportable range, analytical specificity, and stability.
RESULTS
The intra- and inter-assay imprecision was 0%-1.4% and 2.4%-5.1%, respectively. The lower limit of quantitation was 0.3 µg/mL, and the assay was linear up to 10.0 µg/mL. A spike recovery study, using reference standard material, showed recovery was 93.5%--101.3% across the analytical measurement range. Method comparison with a reference laboratory method demonstrated equivalent performance with a slope of 1.01, intercept of 0.09, and correlation coefficient of 0.9948.
CONCLUSION
This assay provides a simple and accurate method for monitoring free carbamazepine with a fast turnaround time.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.