Effects of using auto flow factor function on hemodialysis efficacy in regular hemodialysis patients.
Аннотация
BACKGROUND
Recent HD systems have moved from pre-defined dialysate flow rates (e.g. 300, 500, 800 mL/min) to an individual setting, or an automatic adjustment with a fixed factor of the dialysate flow to the blood flow. Both allows more flexibility of choosing the appropriate dialysate flow as an optimal adjustment to the current blood flow.
AIM OF THE WORK: To study the possible effects of using autoflow factor function on hemodialysis (HD) efficacy in regular hemodialysis patients.
METHODS
A prospective cross section study included 55 ESRD patients on hemodialysis and conducted in hemodialysis unit in Menoufia university hospital.
RESULTS
There was no statistically significant difference between the dialysate flow rate 500 mL/min and auto Flow factor regarding the Kt/V (p = 0.068).
CONCLUSION
The present study demonstrates that the use of the Auto Flow Factor Function does not result in a statistically significant difference in dialysis efficacy compared to a fixed dialysate flow rate of 500 mL/min. Given the potential for dialysate and energy savings, the implementation of Auto Flow may represent a step toward environmentally sustainable dialysis without compromising patient care.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.