Electrical Cardiometry to Monitor Cardiac Output in Preterm Infants with Patent Ductus Arteriosus: A Comparison with Echocardiography.

Hsu Kai-Hsiang, Wu Tai-Wei, Wu I-Hsyuan, Lai Mei-Yin, Hsu Shih-Yun, Huang Hsiao-Wen, Mok Tze-Yee, Lien Reyin

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Electrical cardiometry (EC) is an impedance-based monitoring that provides noninvasive cardiac output (CO) assessment. Through comparison to transthoracic echocardiography (Echo), the accuracy of EC has been verified. However, left-to-right patent ductus arteriosus (PDA) shunting is a concern because PDA shunts aortic flow to the pulmonary artery and may interfere with EC in measuring CO.

OBJECTIVE

To determine the agreement between EC and Echo in preterm infants with a hemodynamically significant PDA (hsPDA).

METHODS

We reviewed our hemodynamic database in which simultaneous CO measurements by Echo and EC (Aesculon®) were recorded. Preterm infants with left-to-right shunting hsPDA were enrolled.

RESULTS

A total of 105 paired measurements in 36 preterm infants were compared. Infants' median (range) age and weight at measurement were 27+2 weeks (24+0-33+1) and 1,015 g (518-1,880), with mean (95% CI) ductal diameter 2.11 mm (1.99-2.22) or 2.15 mm/kg (2.00-2.30). Mean COEC and COEcho were 252 ± 32 and 258 ± 45 mL/kg/min, respectively, which demonstrated a moderate correlation and without a significant between-measurement difference. Bland-Altman analysis showed a bias, limits of agreement, and error percentage of -5.3 mL/kg/min, -78.3 to 67.7 mL/kg/min, and 28.6%, respectively. There was a trend of increased bias and error percentage of infants with high CO ≥280 mL/kg/min and supported with high-frequency ventilator.

CONCLUSIONS

EC and Echo have a wide but clinically acceptable agreement in measuring CO in preterm infants with hsPDA. However, for infants with high CO or ventilated by high-frequency ventilation, interpretation of COEC should be approached with caution.

Для цитирования: Hsu Kai-Hsiang, Wu Tai-Wei, Wu I-Hsyuan, Lai Mei-Yin, Hsu Shih-Yun, Huang Hsiao-Wen, Mok Tze-Yee, Lien Reyin. Electrical Cardiometry to Monitor Cardiac Output in Preterm Infants with Patent Ductus Arteriosus: A Comparison with Echocardiography.. Neonatology. 2017. DOI: 10.1159/000475774.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.