OBJECTIVE

The use of cadmium-zinc-telluride-based scanners may increase the clinical feasibility of simultaneous dual-isotope imaging. In the current study, we sought to investigate a potential acquisition time in simultaneous Tc-tetrofosmin/I-β-methyl-p-iodophenyl pentadecanoic acid dual-isotope imaging using a Discovery NM/CT 670 cadmium-zinc-telluride.

METHODS

Simultaneous Tc-tetrofosmin/I-β-methyl-p-iodophenyl pentadecanoic acid dual-isotope imaging was performed in 29 patients who had undergone primary percutaneous coronary intervention for acute myocardial infarction. Referenced images with an acquisition time of 65 s/view (16.25 min) were reframed to produce images with acquisition times of 33, 16, and 8 s/view. The values for the quantitative-gated single-photon emission computed tomography (SPECT) and the quantitative perfusion SPECT were compared.

RESULTS

The quantitative-gated SPECT values for images with 33, 16, and 8 s/views showed good consistency with those for 65 s/view (the lower 95% confidence intervals for the intraclass correlation were ≥0.80). The quantitative perfusion SPECT values for Tc-tetrofosmin images with 33, 16, and 8 s/views also showed good consistency with those for 65 s/view; however, the quantitative perfusion SPECT values for I-β-methyl-p-iodophenyl pentadecanoic acid images with an acquisition time of 8 s/view were not consistent with the reference acquisition time of 65 s/view.

CONCLUSIONS

The quantitative-gated SPECT and quantitative perfusion SPECT values obtained from images with shorter acquisition times correlated with the values obtained from images with a reference acquisition time of 65 s/view; however, tracer-specific predisposition should be considered. These findings suggest that it is possible to reduce acquisition time when performing simultaneous Tc-tetrofosmin/I-β-methyl-p-iodophenyl pentadecanoic acid dual-tracer imaging with the novel cadmium-zinc-telluride scanner.

Для цитирования: Yamada Yoshihiro, Nakano Shintaro, Gatate Yodo, Sugi Keiki, Okano Nanami, Muramatsu Toshihiro, Nishimura Shigeyuki, Kuji Ichiei, Fukushima Kenji, Matsunari Ichiro. Shortened acquisition time in simultaneous 99mTc-tetrofosmin and 123I-β-methyl-p-iodophenyl pentadecanoic acid dual-tracer imaging with cadmium-zinc-telluride detectors in patients undergoing primary coronary intervention for acute myocardial infarction.. Nuclear medicine communications. 2019. DOI: 10.1097/mnm.0000000000001084.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.