Медицинская библиотека · Biochemia medica · 2025

The analytical impact of extracellular vesicles PSA on different commercial total PSA measurement methods.

Moreno Ana, Sandúa Amaia, Ferrer-Costa Roser, Jacobs-Cacha Conxita, Varo Nerea, Ancizu-Marckert Javier, Robles Jose Enrique, Pérez Gracia Jose Luis, Alegre Estibaliz, González Álvaro

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Prostate-specific antigen (PSA) can circulate bound to extracellular vesicles (EVs) and its measurement (ev-PSA) can be useful in prostate cancer. Although not designed with that purpose, total PSA assays react with ev-PSA. We evaluated the analytical performance of several total PSA assays in ev-PSA quantification and the impact of ev-PSA on total PSA measurement.

MATERIALS AND

METHODS

Extracellular vesicles were isolated from 83 serum samples from prostate cancer patients by size exclusion chromatography or ultracentrifugation. PSA was quantified in serum, EVs, International Standard for PSA 17/100 from the World Health Organization (WHO IS 17/100) and exosomes from lymph node carcinoma of the prostate (LNCaP) cell line, using commercial immunoassays (Elecsys, Atellica, Immulite, Liaison and Kryptor).

RESULTS

Nanoparticle tracking analysis showed that the WHO IS 17/100 contains significantly less EVs than serum (P < 0.001). The sensitivity to detect ev-PSA followed this order: Elecsys ~ Atellica > Immulite > Liaison > Kryptor. Ev-PSA could be detected in all serum samples with Elecsys and Atellica, but not with Immulite (87.8%), Liaison (58.5%) or Kryptor (48.8%). Bland-Altman analysis showed a proportional bias in ev-PSA quantification between Elecsys and other methods. Addition of ev-PSA to serum samples caused a proportional bias in PSA measurement between Elecsys and Immulite methods, with a relationship (r2 = 0.99; P < 0.001) between ev-PSA and the difference in total PSA concentration between both methods.

CONCLUSIONS

While ev-PSA can be measured using commercial kits, notable differences exist between methods, which could lead to potential discrepancies in serum total PSA results across various assays.

Для цитирования: Moreno Ana, Sandúa Amaia, Ferrer-Costa Roser, Jacobs-Cacha Conxita, Varo Nerea, Ancizu-Marckert Javier, Robles Jose Enrique, Pérez Gracia Jose Luis, Alegre Estibaliz, González Álvaro. The analytical impact of extracellular vesicles PSA on different commercial total PSA measurement methods.. Biochemia medica. 2025. DOI: 10.11613/bm.2026.010703.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.