Short term evaluation of respiratory effort by premature infants supported with bubble nasal continuous airway pressure using Seattle-PAP and a standard bubble device.

Welty Stephen E, Rusin Craig G, Stanberry Larissa I, Mandy George T, Gest Alfred L, Ford Jeremy M, Backes Carl H, Richardson C Peter, Howard Christopher R, Hansen Thomas N, Smith Charles V

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Almost one million prematurely born infants die annually from respiratory insufficiency, predominantly in countries with limited access to respiratory support for neonates. The primary hypothesis tested in the present study was that a modified device for bubble nasal continuous positive airway pressure (Bn-CPAP) would provide lower work of spontaneous breathing, estimated by esophageal pressure-rate products.

METHODS

Infants born <32 weeks gestation and stable on Bn-CPAP with FiO2 <0.30 were studied within 72 h following delivery. Esophageal pressures during spontaneous breathing were measured during 2 h on standard Bn-CPAP, then 2 h with Bn-CPAP using a modified bubble device presently termed Seattle-PAP, which produces a different pattern of pressure fluctuations and which provided greater respiratory support in preclinical studies, then 2 h on standard Bn-CPAP.

RESULTS

All 40 infants enrolled completed the study and follow-up through 36 wks post menstrual age or hospital discharge, whichever came first. No infants were on supplemental oxygen at completion of follow-up. No infants developed pneumothoraces or nasal trauma, and no adverse events attributed to the study were observed. Pressure-rate products on the two devices were not different, but effort of breathing, assessed by areas under esophageal pressure-time curves, was lower with Seattle-PAP than with standard Bn-CPAP.

CONCLUSION

Use of Seattle-PAP to implement Bn-CPAP lowers the effort of breathing exerted even by relatively healthy spontaneously breathing premature neonates. Whether the lower effort of breathing observed with Seattle-PAP translates to improvements in neonatal mortality or morbidity will need to be determined by studies in appropriate patient populations.

Для цитирования: Welty Stephen E, Rusin Craig G, Stanberry Larissa I, Mandy George T, Gest Alfred L, Ford Jeremy M, Backes Carl H, Richardson C Peter, Howard Christopher R, Hansen Thomas N, Smith Charles V. Short term evaluation of respiratory effort by premature infants supported with bubble nasal continuous airway pressure using Seattle-PAP and a standard bubble device.. PloS one. 2018. DOI: 10.1371/journal.pone.0193807.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.