A peripherally inserted central catheter (PICC) is commonly used for fluid infusion in patients. However, rupture is one of the most serious complications associated with PICC placement. We investigated an 81-year-old patient who experienced a catheter break nearly 11 months after the placement of PICC, followed by a catheter break during catheter capture that was removed after accessing the superior vena cava with a catcher. This case suggests that silicone-based PICCs are fragile and have a high risk of spontaneous dislocation. Therefore, they should be replaced with polyurethane-based PICCs.