Validation of the OMRON M7 (HEM-780-E) blood pressure measuring device in a population requiring large cuff use according to the International Protocol of the European Society of Hypertension.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
El Feghali Ramzi N, Topouchian Jirar A, Pannier Bruno M, El Assaad Hiba A, Asmar Roland G, European Society of Hypertension
Аннотация
BACKGROUND: A high percentage of hypertensive patients present an arm circumference of over 32 cm; the use of a large cuff is therefore recommended. Validation studies are usually performed in the general population using a standard-size cuff. The aim of this study was to assess the accuracy of the Omron M7 device in a population with an arm circumference ranging from 32 to 42 cm.
DESIGN: A validation study was performed according to the International Protocol of the European Society of Hypertension. This protocol is divided into two phases: the first phase is performed on 15 selected participants (45 pairs of blood-pressure measurements); if the device passes this phase, 18 supplementary participants are included (54 pairs of blood-pressure measurements), making a total number of 33 participants (99 pairs of blood-pressure measurements), on whom the analysis is performed.
METHODS: For each participant, four blood-pressure measurements were performed simultaneously by two trained observers, using mercury sphygmomanometers fitted with a Y tube; the measurements alternated with three by the test device. The difference between the blood-pressure value given by the device and that obtained by the two observers (mean of the two observations) was calculated for each measure. The 99 pairs of blood-pressure differences were classified into three categories (<or=5, <or=10 and <or=15 mmHg). The number of differences in each category was compared with the number required by the European Society of Hypertension protocol.
RESULTS: The Omron M7 device passed the first and the second phases of the validation process. The average differences between the two observers were 1.5+/-3.2 and -0.5+/-2.2 mmHg for systolic blood pressure and diastolic blood pressure, and those between the device and the mercury sphygmomanometer were -1.6+/-6.7 for systolic blood pressure and -0.12+/-4.0 mmHg for diastolic blood pressure Readings that differ by less than 5, 10 and 15 mmHg for systolic blood-pressure and diastolic blood-pressure values fulfill the recommendation criteria of the European Society of Hypertension protocol.
CONCLUSIONS: The Omron M7 (HEM-780-E) device fulfilled the validation criteria of the international protocol in a population with an arm circumference ranging from 32 to 42 cm.
DESIGN: A validation study was performed according to the International Protocol of the European Society of Hypertension. This protocol is divided into two phases: the first phase is performed on 15 selected participants (45 pairs of blood-pressure measurements); if the device passes this phase, 18 supplementary participants are included (54 pairs of blood-pressure measurements), making a total number of 33 participants (99 pairs of blood-pressure measurements), on whom the analysis is performed.
METHODS: For each participant, four blood-pressure measurements were performed simultaneously by two trained observers, using mercury sphygmomanometers fitted with a Y tube; the measurements alternated with three by the test device. The difference between the blood-pressure value given by the device and that obtained by the two observers (mean of the two observations) was calculated for each measure. The 99 pairs of blood-pressure differences were classified into three categories (<or=5, <or=10 and <or=15 mmHg). The number of differences in each category was compared with the number required by the European Society of Hypertension protocol.
RESULTS: The Omron M7 device passed the first and the second phases of the validation process. The average differences between the two observers were 1.5+/-3.2 and -0.5+/-2.2 mmHg for systolic blood pressure and diastolic blood pressure, and those between the device and the mercury sphygmomanometer were -1.6+/-6.7 for systolic blood pressure and -0.12+/-4.0 mmHg for diastolic blood pressure Readings that differ by less than 5, 10 and 15 mmHg for systolic blood-pressure and diastolic blood-pressure values fulfill the recommendation criteria of the European Society of Hypertension protocol.
CONCLUSIONS: The Omron M7 (HEM-780-E) device fulfilled the validation criteria of the international protocol in a population with an arm circumference ranging from 32 to 42 cm.