BACKGROUND

This study examined the degree of hemolysis during vacuum-assisted venous drainage at different negative pressures to identify an adequate negative pressure that provides effective venous drainage without significant hemolysis in open-heart surgery in children weighing less than 10 kg.

METHODS

Patients weighing less than 10 kg who underwent surgery for ventricular septal defect or atrial septal defect from 2011 to 2014 were enrolled. We used one of four negative pressures (20, 30, 40, or 60 mm Hg) for each patient. We measured haptoglobin, plasma hemoglobin, aspartate aminotransferase, and lactate dehydrogenase levels in the patients' blood three times perioperatively and determined the potential correlation between the change in each parameter with the level of negative pressure.

RESULTS

Forty-six patients were enrolled in this study (mean age: 7.1 ± 7.0 months, mean body weight: 6.1 ± 1.8 kg). There were no significant differences according to the degree of negative pressure with respect to patient age, body weight, cardiopulmonary bypass (CPB) time, aorta cross-clamping time, blood flow during CPB, or lowest body temperature. All parameters that we measured reflected progression of hemolysis during CPB; however, the degree of change in the parameters did not correlate with negative pressure.

CONCLUSION

In pediatric patients weighing less than 10 kg, the change in the degree of hemolysis did not differ with the amount of negative pressure. We may apply negative pressures up to 60 mm Hg without increasing the risk of hemolysis, with almost same the level of hemolysis using negative pressures of 20, 30, and 40 mm Hg for effective venous drainage and an ideal operative field during open-heart surgery.

Для цитирования: Kwak Jae Gun, Lee Jinkwon, Park Minkyoung, Seo Yu-Jin, Lee Chang-Ha. Hemolysis During Open-Heart Surgery With Vacuum-Assisted Venous Drainage at Different Negative Pressures in Pediatric Patients Weighing Less Than 10 kilograms.. World journal for pediatric & congenital heart surgery. 2017. DOI: 10.1177/2150135116681734.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.