DOI 10.3109/02841850903321625

Медицинская библиотека · Acta radiologica (Stockholm, Sweden : 1987) · 2010

Detection of breast cancer by soft-copy reading of digital mammograms: comparison between a routine image-processing parameter and high-contrast parameters.

Kamitani Takeshi, Yabuuchi Hidetake, Soeda Hiroyasu, Matsuo Yoshio, Okafuji Takashi, Sakai Shuji, Setoguchi Taro, Hatakenaka Masamitsu, Ishii Nobuhide, Honda Hiroshi

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Recent studies have reported the clinical usefulness of the soft-copy reading of mammograms. However, image-processing parameters for soft-copy reading of digital mammograms have not been established.

PURPOSE: To compare observer performance in detecting breast cancer by soft-copy reading of digital mammograms using a routine image-processing parameter versus each of several high-contrast parameters.

MATERIAL AND

METHODS

The mammograms of 154 breasts, including 48 abnormal breasts with breast cancer and 106 normal breasts, were examined. Cancers were classified into 34 mass-dominant cancers, 11 microcalcification-dominant cancers, two cancers showing only architectural distortion, and one cancer without abnormal findings. All mammograms were performed using a computed radiography (CR) system. Each image was processed using GA (1.2), which was the contrast parameter recommended by the manufacturer for hard-copy film, GA (1.4), GA (1.6), and GA (1.8). These images were displayed on 5-megapixel (M) liquid-crystal display monitors. Five experienced radiologists classified them into BI-RADS category 1-2 or 3-5, and were also asked to rate the images on a scale of 0 to 100 for the likelihood of the presence of masses and microcalcifications in each breast.

RESULTS

In mass-dominant cancers of dense breast tissue, the mean sensitivities of GA (1.2), GA (1.4), GA (1.6), and GA (1.8) were 32.7, 38.2, 36.4, and 40.0, and the A(Z) values were 0.67, 0.73, 0.71, and 0.73, respectively; in microcalcification-dominant cancers, the mean sensitivities were 80.0, 74.5, 80.0, and 78.2, respectively; however, there were no significant differences among them.

CONCLUSION

High-contrast parameters tended to show relatively high sensitivity and A(Z) values in the detection of masses in dense breast tissue, but relatively low sensitivity for microcalcifications.

Для цитирования: Kamitani Takeshi, Yabuuchi Hidetake, Soeda Hiroyasu, Matsuo Yoshio, Okafuji Takashi, Sakai Shuji, Setoguchi Taro, Hatakenaka Masamitsu, Ishii Nobuhide, Honda Hiroshi. Detection of breast cancer by soft-copy reading of digital mammograms: comparison between a routine image-processing parameter and high-contrast parameters.. Acta radiologica (Stockholm, Sweden : 1987). 2010. DOI: 10.3109/02841850903321625.