Use of Chest Wall Electromyography to Detect Respiratory Effort during Polysomnography.

Berry Richard B, Ryals Scott, Girdhar Ankur, Wagner Mary H

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

OBJECTIVES

To evaluate the ability of chest wall EMG (CW-EMG) using surface electrodes to classify apneas as obstructive, mixed, or central compared to classification using dual channel uncalibrated respiratory inductance plethysmography (RIP).

METHODS

CW-EMG was recorded from electrodes in the eighth intercostal space at the right mid-axillary line. Consecutive adult clinical sleep studies were retrospectively reviewed, and the first 60 studies with at least 10 obstructive and 10 mixed or central apneas and technically adequate tracings were selected. Four obstructive and six central or mixed apneas (as classified by previous clinical scoring) were randomly selected. A blinded experienced scorer classified the apneas on the basis of tracings showing either RIP channels or the CW-EMG channel. The agreement using the two classification methods was determined by kappa analysis and intraclass correlation.

RESULTS

The percentage agreement was 89.5%, the kappa statistic was 0.83 (95% confidence interval 0.79 to 0.87), and the intraclass correlation was 0.83, showing good agreement. Of the 249 apneas classified as central by RIP, 26 were classified as obstructive (10.4%) and 7 as mixed (2.8%) by CW-EMG. Of the 229 events classified as central by CW-EMG, 7 (3.1%) were classified as obstructive and 6 (2.6%) as mixed by RIP.

CONCLUSIONS

Monitoring CW-EMG may provide a clinically useful method of detection of respiratory effort when used with RIP and can prevent false classification of apneas as central. RIP can rarely detect respiratory effort not easily discernible by CW-EMG and the combination of the two methods is more likely to avoid apnea misclassification.

Для цитирования: Berry Richard B, Ryals Scott, Girdhar Ankur, Wagner Mary H. Use of Chest Wall Electromyography to Detect Respiratory Effort during Polysomnography.. Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine. 2016. DOI: 10.5664/jcsm.6122.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.