Performance of digital morphology analyzer Vision Pro on white blood cell differentials.
Аннотация
OBJECTIVES
Vision Pro (West Medica, Perchtoldsdorf, Austria) is a recently developed digital morphology analyzer. We evaluated the performance of Vision Pro on white blood cell (WBC) differentials.
METHODS
In a total of 200 peripheral blood smear samples (100 normal and 100 abnormal samples), WBC preclassification and reclassification by Vision Pro were evaluated and compared with manual WBC count, according to the Clinical and Laboratory Standards Institute guidelines (H20-A2).
RESULTS
The overall sensitivity was high for normal WBCs and nRBCs (80.1-98.0%). The overall specificity and overall efficiency were high for all cell classes (98.1-100.0% and 97.7-99.9%, respectively). The absolute values of mean differences between Vision Pro and manual count ranged from 0.01 to 1.31. In leukopenic samples, those values ranged from 0.09 to 2.01. For normal WBCs, Vision Pro preclassification and manual count showed moderate or high correlations (r=0.52-0.88) except for basophils (r=0.34); after reclassification, the correlation between Vision Pro and manual count was improved (r=0.36-0.90).
CONCLUSIONS
This is the first study that evaluated the performance of Vision Pro on WBC differentials. Vision Pro showed reliable analytical performance on WBC differentials with improvement after reclassification. Vision Pro could help improve laboratory workflow.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.