Accuracy of the smartphone blood pressure measurement solution OptiBP to track blood pressure changes in pregnant women.

Almeida Pedro, Cuénoud Alexia, Hoang Harry, Othenin-Girard Alexandra, Salhi Nadia, Köthe Andreas, Christen Urvan, Schoettker Patrick

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Hypertensive disorders present significant morbidity and mortality during pregnancy. Although ambulatory blood pressure measurement remains the standard of care for normotensive women, self-monitoring at home is increasingly prevalent. The widespread use of smartphones worldwide has sparked interest in mobile applications that leverage the built-in hardware for blood pressure estimation, yet few trials have assessed their accuracy.

METHODS

This prospective, longitudinal and monocentric study evaluated the accuracy of the OptiBP algorithm against standard oscillometric blood pressure measurements in a sample of pregnant women. Patients scheduled for elective caesarean sections were enrolled during the preoperative anesthesia consultations. Paired blood pressure measurements using OptiBP and the reference method were obtained at multiple time-points in late pregnancy and the postpartum period. Agreement between methods was assessed using the AAMI/ESH/ISO 81060-2:2018 standard thresholds of 5 ± 8 mmHg for mean ± standard deviation of the error (criterion 1) and patient-specific standard deviation of the mean error (criterion 2) and represented graphically by Bland-Altman scatterplots.

RESULTS

Forty-eight women were enrolled of which 32 completed the protocol, yielding 338 total valid measurement pairs. Mean and standard deviation of the error were -1.78 ± 7.94 and 1.19 ± 7.59, and the patient-specific standard deviation of the mean error was 4.68 and 4.52, for SBP and DBP, respectively.

CONCLUSION

Compared with blood pressure measurements taken with an oscillometric device, OptiBP's blood pressure estimates meet the AAMI/ESH/ISO 81060-2:2018 criteria.

Для цитирования: Almeida Pedro, Cuénoud Alexia, Hoang Harry, Othenin-Girard Alexandra, Salhi Nadia, Köthe Andreas, Christen Urvan, Schoettker Patrick. Accuracy of the smartphone blood pressure measurement solution OptiBP to track blood pressure changes in pregnant women.. Journal of hypertension. 2025. DOI: 10.1097/hjh.0000000000003956.

Предметные рубрики MeSH