Performance comparison of platelet function analyzers in cardiology patients: VerifyNow and Anysis-200 aspirin assays.

Chung Young Hak, Lee Kyung Ah, Cho Minhee, Shin Sehyun, Lee Byoung Kwon

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Analysis of responsiveness to antiplatelet therapy is crucial in the management of patients with cardiovascular diseases.

OBJECTIVE

This study aimed to evaluate a new platelet function analysis system (Anysis-200) and to compare it with VerifyNow (Accumetrics, San Diego, CA, USA) in cardiology patients.

METHODS

Overall, 125 citrated blood samples were collected from 85 cardiology patients referred for platelet function testing. In Anysis-200, platelet function was measured as blood migration distance (MD) until clogging of flow passage, which is comparable to aspirin resistance units obtained using VerifyNow. The two devices were simultaneously used and compared.

RESULTS

The MDs before and after taking aspirin were 175±51 and 247±27 mm, respectively (p < 0.0001). Compared with VerifyNow (reference), the sensitivity and specificity of Anysis-200 was 91.5% and 75.5%, respectively (area under the curve, 0.829). Further, the true positive rate in patients newly taking aspirin was 85% for VerifyNow and 92.5% for Anysis-200, respectively. The Cohen's kappa coefficient between the two devices was 0.682, indicating a relatively high agreement.

CONCLUSIONS

Anysis-200, a novel system for assessing platelet aggregation, has accuracy and precision equivalent to that of, and significant agreement with, VerifyNow. Anysis-200 may be useful in screening patients with abnormal platelet reactivity and aspirin nonresponsiveness.

Для цитирования: Chung Young Hak, Lee Kyung Ah, Cho Minhee, Shin Sehyun, Lee Byoung Kwon. Performance comparison of platelet function analyzers in cardiology patients: VerifyNow and Anysis-200 aspirin assays.. Clinical hemorheology and microcirculation. 2020. DOI: 10.3233/ch-200822.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.