Incomplete Expansion of Chimney Stent Graft during Chimney-Thoracic Endovascular Aneurysm Repair.
Клинический случайСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Katsargyris Athanasios, Spinelli Domenico, Oikonomou Kyriakos, Mufty Hozan, Verhoeven Eric Lg
Аннотация
BACKGROUND: To report a technical complication during a chimney-thoracic endovascular aneurysm repair (Ch-TEVAR) procedure.
METHODS: A 77-year-old female patient underwent Ch-TEVAR for a symptomatic thoracic aortic aneurysm with a short and angulated proximal neck. Deployment of the aortic stent graft over the left subclavian artery (LSA) was followed by placement of a balloon-expandable covered stent as a chimney for the LSA.
RESULTS: Chimney deployment failed due to balloon perforation. Additional attempts with several balloons all failed, albeit gaining some additional expansion of the chimney stent graft. The technical complication was solved by deploying a self-expanding bare stent inside the chimney stent. This enabled additional ballooning and further expansion of the chimney stent graft leading to an acceptable end result.
CONCLUSIONS: Ch-TEVAR with a balloon-expandable chimney stent graft can be complicated by incomplete chimney deployment due to balloon perforation. Relining of the chimney stent graft with a self-expanding stent improved the deployment with acceptable patency.
METHODS: A 77-year-old female patient underwent Ch-TEVAR for a symptomatic thoracic aortic aneurysm with a short and angulated proximal neck. Deployment of the aortic stent graft over the left subclavian artery (LSA) was followed by placement of a balloon-expandable covered stent as a chimney for the LSA.
RESULTS: Chimney deployment failed due to balloon perforation. Additional attempts with several balloons all failed, albeit gaining some additional expansion of the chimney stent graft. The technical complication was solved by deploying a self-expanding bare stent inside the chimney stent. This enabled additional ballooning and further expansion of the chimney stent graft leading to an acceptable end result.