MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ando.2024.101677

Медицинская библиотека · Annales d'endocrinologie · 2024

Evaluation of the analytic performance and macroprolactin sensitivity of a new prolactin immunoassay.

David Guillaume, Perrin Pauline, Sergeant Camille, Raverot Gérald, Raverot Véronique

СтатьяОценочное исследованиеСравнительное исследованиеПеревод в очереди

Аннотация

PURPOSE: Prolactin measurement is essential in endocrine diagnostics. Challenges such as the hook effect and reactivity to macroprolactin, which varies according to the reagent, complicate accurate measurement. The present study evaluated a newly marketed reagent to detect prolactin, IDS Prolactin, comparing it to an established reagent, Roche Elecsys Prolactin, assessing its behavior toward macroprolactin and polyethylene glycol (PEG) treatment, and establishing reference intervals.

METHODS

The IDS Prolactin and Roche Elecsys Prolactin assays were compared using 44 samples containing macroprolactin confirmed on gel filtration chromatography (macroprolactin: BBPRL) and 104 samples for which the diagnosis of macroprolactin was excluded (monomeric prolactin: MNPRL). Analytic performance of the IDS Prolactin assay was also assessed.

RESULTS

The new reagent showed satisfactory analytic performance, meeting EFLM standards for repeatability and intermediate imprecision. Comparison between the two methods found robust correlation for monomeric samples (y=1.060x-18.28; r2=0.993). Compared to the Roche assay, which is particularly low in its reaction to macroprolactin, the IDS assay displayed a higher level of detection. PEG precipitation effectively separated monomeric and macroprolactin samples when a cut-off of 65% recovery was used or at the threshold of 444 mIU/L (20.9μg/L) for post-PEG monomeric prolactin upper limit of normal. Reference intervals were established for women, with ROC curve analysis demonstrating high sensitivity and specificity.

CONCLUSION

The IDS Prolactin assay showed excellent analytic performance and satisfactory characteristics on macroprolactinemic samples.

Для цитирования: David Guillaume, Perrin Pauline, Sergeant Camille, Raverot Gérald, Raverot Véronique. Evaluation of the analytic performance and macroprolactin sensitivity of a new prolactin immunoassay.. Annales d'endocrinologie. 2024. DOI: 10.1016/j.ando.2024.101677.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.