New technique for C1-C2 fixation.
Аннотация
BACKGROUND
There are several techniques for treating atlantoaxial instability, including the Magerl transarticular screw fixation and the Harms/Goel C1-C2 screw rod techniques. Here, we present a novel technique utilizing a polyaxial screw rod system and a combination of C1 lateral mass and C1-C2 transarticular screws.
METHODS
We retrospectively reviewed 14 patients (7 women, 7 men; mean age 62) who underwent surgery for type II odontoid fractures (n = 7), pseudarthrosis after anterior odontoid screw placement (n = 3), Os odontoideum (n = 2), atlantoaxial instability after C3-C5 fusion (n = 1), and craniovertebral rheumatoid arthritis (n = 1). Ten patients underwent posterior C1-C2 fixation, three patients with osteoporosis had C1-C4 fixation, and one patient had C1-Th1 fixation. The mean follow-up time was 22 months.
RESULTS
Intraoperatively, there were no complications (e.g., vertebral artery, nerve root, or spinal cord injury). Postoperative imaging showed no screw malpositioning, and no screw loosening, fracture, or bone absorption around the screws. Furthermore, all patients exhibited postoperative improvement in neck pain.
CONCLUSIONS
C1 lateral mass and C1-C2 transarticular polyaxial screw rod fixation techniques were effective in achieving immediate rigid immobilization of the C1-C2 motion segment.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.