DOI 10.1148/radiol.2521081621

Медицинская библиотека · Radiology · 2009

Breast percent density: estimation on digital mammograms and central tomosynthesis projections.

Bakic Predrag R, Carton Ann-Katherine, Kontos Despina, Zhang Cuiping, Troxel Andrea B, Maidment Andrew D A

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To evaluate inter- and intrareader agreement in breast percent density (PD) estimation on clinical digital mammograms and central digital breast tomosynthesis (DBT) projection images.

MATERIALS AND

METHODS

This HIPAA-compliant study had institutional review board approval; all patients provided informed consent. Breast PD estimation was performed on the basis of anonymized digital mammograms and central DBT projections in 39 women (mean age, 51 years; range, 31-80 years). All women had recently detected abnormalities or biopsy-proved cancers. PD was estimated by three experienced readers on the mediolateral oblique views of the contralateral breasts by using software; each reader repeated the estimation after 2 months. Spearman correlations of inter- and intrareader and intermodality PD estimates, as well as kappa statistics between categoric PD estimates, were computed.

RESULTS

High correlation (rho = 0.91) was observed between PD estimates on digital mammograms and those on central DBT projections, averaged over all estimations; the corresponding kappa coefficient (0.79) indicated substantial agreement. Mean interreader agreement for PD estimation on central DBT projections (rho = 0.85 +/- 0.05 [standard deviation]) was significantly higher (P < .01) than that for PD estimation on digital mammograms (rho = 0.75 +/- 0.05); the corresponding kappa coefficients indicated substantial (kappa = 0.65 +/- 0.12) and moderate (kappa = 0.55 +/- 0.14) agreement for central DBT projections and digital mammograms, respectively.

CONCLUSION

High correlation between PD estimates on digital mammograms and those on central DBT projections suggests the latter could be used until a method for PD estimation based on three-dimensional reconstructed images is introduced. Moreover, clinical PD estimation is possible with reduced radiation dose, as each DBT projection was acquired by using about 22% of the dose for a single mammographic projection.

Для цитирования: Bakic Predrag R, Carton Ann-Katherine, Kontos Despina, Zhang Cuiping, Troxel Andrea B, Maidment Andrew D A. Breast percent density: estimation on digital mammograms and central tomosynthesis projections.. Radiology. 2009. DOI: 10.1148/radiol.2521081621.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.