m+Medelevator.RU
← Поиск публикаций
Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2016

Initial clinical experience with the Medtronic Micro Vascular Plug™ in transcatheter occlusion of PDAs in extremely premature infants.

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Sathanandam Shyam, Justino Henri, Waller B Rush, Radtke Wolfgang, Qureshi Athar M

Аннотация

OBJECTIVES: To describe the early multicenter, clinical experience with the Medtronic Micro Vascular Plug™ (MVP) for the occlusion of patent ductus arteriosus (PDA) in premature infants.

BACKGROUND: The MVP is a large diameter plug that can be delivered through a microcatheter for occlusion of abnormal blood vessels.

METHODS: A Retrospective review of PDA embolization procedures performed in two centers using the MVP was performed.

RESULTS: Fifteen premature infants underwent attempted PDA occlusion using the MVP. The gestational age and birth weight were 25.6 ± 2.5 weeks and 735 ± 251 g, respectively. The median weight and age at the time of the procedure were 1,210 g (700-3,500 g) and 4.5 weeks (2-12 weeks), respectively. Median procedure and fluoroscopy times were 45 and 6.5 min, respectively. The median radiation and contrast doses were 19.7 mGy and 2.4 mL/kg, respectively. Antegrade occlusion was successfully achieved in 13 patients <2 kg with only femoral venous access aided by echo guidance. The two patients >2 kg had arterial access and attempted retrograde occlusion; one of which was unsuccessful due to the PDA being short and wide. Complete closure was observed in 13 of 14 successful procedures (93%), with one patient having a small residual shunt that was not seen on follow-up. There were no complications related to the procedure or noted during follow-up (Median 11 months).

CONCLUSIONS: The MVP is a new, large-diameter vascular embolization device that may be useful for the occlusion of PDA in extremely small, premature infants. © 2016 Wiley Periodicals, Inc.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты