DOI 10.1002/ccd.26878

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2016

Initial clinical experience with the Medtronic Micro Vascular Plug™ in transcatheter occlusion of PDAs in extremely premature infants.

Sathanandam Shyam, Justino Henri, Waller B Rush, Radtke Wolfgang, Qureshi Athar M

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

To describe the early multicenter, clinical experience with the Medtronic Micro Vascular Plug™ (MVP) for the occlusion of patent ductus arteriosus (PDA) in premature infants.

BACKGROUND

The MVP is a large diameter plug that can be delivered through a microcatheter for occlusion of abnormal blood vessels.

METHODS

A Retrospective review of PDA embolization procedures performed in two centers using the MVP was performed.

RESULTS

Fifteen premature infants underwent attempted PDA occlusion using the MVP. The gestational age and birth weight were 25.6 ± 2.5 weeks and 735 ± 251 g, respectively. The median weight and age at the time of the procedure were 1,210 g (700-3,500 g) and 4.5 weeks (2-12 weeks), respectively. Median procedure and fluoroscopy times were 45 and 6.5 min, respectively. The median radiation and contrast doses were 19.7 mGy and 2.4 mL/kg, respectively. Antegrade occlusion was successfully achieved in 13 patients <2 kg with only femoral venous access aided by echo guidance. The two patients >2 kg had arterial access and attempted retrograde occlusion; one of which was unsuccessful due to the PDA being short and wide. Complete closure was observed in 13 of 14 successful procedures (93%), with one patient having a small residual shunt that was not seen on follow-up. There were no complications related to the procedure or noted during follow-up (Median 11 months).

CONCLUSIONS

The MVP is a new, large-diameter vascular embolization device that may be useful for the occlusion of PDA in extremely small, premature infants. © 2016 Wiley Periodicals, Inc.

Для цитирования: Sathanandam Shyam, Justino Henri, Waller B Rush, Radtke Wolfgang, Qureshi Athar M. Initial clinical experience with the Medtronic Micro Vascular Plug™ in transcatheter occlusion of PDAs in extremely premature infants.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2016. DOI: 10.1002/ccd.26878.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.