Safety and efficacy of a temperature-controlled ablation system for ventricular tachycardia: Results from the TRAC-VT study.

Kautzner Josef, Moreno Javier, Tondo Claudio, Anselme Frédéric, Burrell James, Becker Daniel, Peichl Petr, Patchett Ian, Dhanjal Tarvinder

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Catheter ablation using radiofrequency (RF) energy is an established treatment for ventricular tachycardia (VT). Tissue temperature is a key determinant of successful lesion creation, and yet, it is difficult to measure during conventional RF ablation because of the cooling effect of high-flow rate saline irrigation. The TRAC-VT study evaluated the safety and efficacy of a novel irrigated RF ablation system modulating power based on real-time tissue temperature.

METHODS

Patients with sustained monomorphic VT and structural heart disease (SHD) were enrolled. Catheter ablation was performed in temperature-control mode (irrigation 8 ml/min, temperature set-points 55 or 60 °C, and power output ≤ 50 W), with RF applications for ≤ 45 s. The primary safety endpoint was a composite of cardiovascular-specific serious procedure-related adverse events within 30 days post-ablation. The primary effectiveness endpoint was acute success (i.e., non-inducibility of all clinically relevant VTs).

RESULTS

Thirty-eight patients were enrolled with monomorphic VT (age 68 ± 12 years and 84% male), with an average of 1.7 ± 1.2 VTs targeted per patient. In total, 41 ± 23 RF applications per patient were delivered. Acute procedural success was 100% (95% CI, 91-100%). No primary safety endpoints were observed. Six-month follow-up was completed in 92% of patients with 81% (95% CI, 65-91%) freedom from sustained or treated VT. A repeat ablation was performed in three patients.

CONCLUSIONS

Ablation of VT in SHD, using a temperature-controlled irrigated RF catheter, was safe and effective with a low rate of VT recurrence at 6 months.

Для цитирования: Kautzner Josef, Moreno Javier, Tondo Claudio, Anselme Frédéric, Burrell James, Becker Daniel, Peichl Petr, Patchett Ian, Dhanjal Tarvinder. Safety and efficacy of a temperature-controlled ablation system for ventricular tachycardia: Results from the TRAC-VT study.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2025. DOI: 10.1007/s10840-025-01995-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.