Novel everting urologic access sheath: decreased axial forces during insertion.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Rubenstein Jonathan N, Garcia Maurice, Camargo Affonso H L A, Joel Andrew B, Stoller Marshall L
Аннотация
BACKGROUND: Advancement of urologic instruments through the genitourinary tract is associated with significant axial forces that likely contribute to patient discomfort, even after injection of a local anesthetic, and may lead to mucosal trauma, postprocedural dysuria and hematuria, and increased susceptibility to infection and strictures. Placing an everting urethral sheath prior to instrumentation may decrease these problems.
MATERIALS AND METHODS: Two 7-cm-long, 5-mm diameter urethral luminal models were created, one with and one without an artificial stricture. We measured the forces generated during advancement of a novel everting access sheath (Cystoglide; Percutaneous Systems, Mountain View, CA) through the models in comparison with a representative cystoscope and a urologic dilator simulating a traditional access sheath.
RESULTS: The mean force generated during advancement of the everting sheath was significantly less than that of both the representative cystoscope (P<0.01) and the traditional access sheath (P<0.01). This held true for the urethral models both with and without an artificial stricture (P<0.01) and with and without lubrication (P<0.01).
CONCLUSIONS: This novel introduction sheath markedly decreased the axial forces applied to an artificial urethral luminal wall. It is possible that the clinical use of this technology will decrease the discomfort and potential complications associated with lower urinary-tract endoscopy.
MATERIALS AND METHODS: Two 7-cm-long, 5-mm diameter urethral luminal models were created, one with and one without an artificial stricture. We measured the forces generated during advancement of a novel everting access sheath (Cystoglide; Percutaneous Systems, Mountain View, CA) through the models in comparison with a representative cystoscope and a urologic dilator simulating a traditional access sheath.
RESULTS: The mean force generated during advancement of the everting sheath was significantly less than that of both the representative cystoscope (P<0.01) and the traditional access sheath (P<0.01). This held true for the urethral models both with and without an artificial stricture (P<0.01) and with and without lubrication (P<0.01).
CONCLUSIONS: This novel introduction sheath markedly decreased the axial forces applied to an artificial urethral luminal wall. It is possible that the clinical use of this technology will decrease the discomfort and potential complications associated with lower urinary-tract endoscopy.