DOI 10.3390/pharmaceutics17070919
Quantification of Total and Unbound Selinexor Concentrations in Human Plasma by a Fully Validated Liquid Chromatography-Tandem Mass Spectrometry Method.
Аннотация
Background/
Objectives
Selinexor is a selective nuclear-export inhibitor approved for hematologic malignancies, characterized by extensive plasma protein binding (>95%). However, a validated analytical method to accurately measure the clinically relevant unbound fraction of selinexor in human plasma has not yet been established. This study aimed to develop a fully validated bioanalytical assay for simultaneous quantification of total and unbound selinexor concentrations in human plasma.
Methods
We established and fully validated an analytical method based on liquid chromatography-tandem mass spectrometry (LC-MS/MS) capable of quantifying total and unbound selinexor concentrations in human plasma. Unbound selinexor was separated using ultrafiltration, and selinexor was efficiently extracted from 50 μL of plasma by liquid-liquid extraction. Chromatographic separation was achieved on a C18 column using an isocratic mobile phase (0.1% formic acid:methanol, 12:88 v/v) with a relatively short runtime of 2.5 min.
Results
Calibration curves showed excellent linearity over a range of 5-2000 ng/mL for total selinexor (r2 ≥ 0.998) and 0.05-20 ng/mL for unbound selinexor (r2 ≥ 0.995). The precision (%CV ≤ 10.35%) and accuracy (92.5-104.3%) for both analytes met the regulatory criteria. This method successfully quantified selinexor in plasma samples from renally impaired patients with multiple myeloma, demonstrating potential inter-individual differences in unbound drug concentrations.
Conclusions
This validated bioanalytical assay enables precise clinical pharmacokinetic assessments in a short runtime using a small plasma volume and, thus, assists in individualized dosing of selinexor, particularly for renally impaired patients with altered protein binding.