Home blood pressure self-monitoring: diagnostic performance in white-coat hypertension.

Bayó Joan, Cos Francesc Xavier, Roca Carme, Dalfó Antoni, Martín-Baranera Maria Montserrat, Albert Botey

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To determine the diagnostic performance of home blood pressure self-monitoring in white-coat hypertension using a 3-day reading program.

MATERIAL AND

METHODS

One hundred and ninety nontreated patients recently diagnosed with mild-moderate hypertension, selected consecutively at four primary healthcare centers in the city of Barcelona, were included. Each patient underwent morning and night home blood pressure self-monitoring with readings in triplicate for three consecutive days, followed by 24-h ambulatory blood pressure monitoring. The normality cut-off point value for home blood pressure self-monitoring and daytime ambulatory blood pressure monitoring was 135/85 mmHg.

RESULTS

Sixty-three patients were diagnosed with white-coat hypertension with home blood pressure self-monitoring (34.8%; 95% confidence interval: 27.9-42.2) and 74 with ambulatory blood pressure monitoring (41.6%; 95% confidence interval: 33.7-48.4). No statistically significant differences were observed between home blood pressure self-monitoring values and those of diurnal ambulatory blood pressure monitoring [137.4 (14.3)/82.1 (8.3) mmHg vs. 134.8 (11.3)/81.3 (9.5) mmHg]. Home blood pressure self-monitoring diagnostic performance parameters were sensitivity 50.0% (95% confidence interval: 38.3-61.7), specificity 75.7% (95% confidence interval: 66.3-83.2), positive and negative predictive values 58.7% (95% confidence interval: 45.6-70.8) and 68.6% (95% confidence interval: 59.4-76.7), respectively, and positive and negative probability coefficients 2.05 and 0.66, respectively. Analysis of different normality cut-off points using a receiver operating characteristic curve failed to produce significant improvement in the diagnostic performance of home blood pressure self-monitoring.

CONCLUSIONS

The diagnostic accuracy of a 3-day home blood pressure self-monitoring reading program in white-coat hypertension was poor. Ambulatory blood pressure monitoring continues to be the test of choice for this indication.

Для цитирования: Bayó Joan, Cos Francesc Xavier, Roca Carme, Dalfó Antoni, Martín-Baranera Maria Montserrat, Albert Botey. Home blood pressure self-monitoring: diagnostic performance in white-coat hypertension.. Blood pressure monitoring. 2006. DOI: 10.1097/01.mbp.0000200479.19046.94.

Предметные рубрики MeSH