DOI 10.1016/j.jhsa.2014.01.038

Медицинская библиотека · The Journal of hand surgery · 2014

Distal radius volar locking plate design and associated vulnerability of the flexor pollicis longus.

Limthongthang Roongsak, Bachoura Abdo, Jacoby Sidney M, Osterman A Lee

СтатьяПеревод в очереди

Аннотация

PURPOSE: Flexor pollicis longus (FPL) tendon rupture is a well-documented complication related to the use of distal radius volar locking plates (VLPs). The final common pathway of flexor tendon rupture appears to involve implants prominent at the watershed line. We hypothesized that significant differences in VLP prominence exist between various plate designs.

METHODS

Ten fresh frozen specimens were dissected to identify the path of the FPL in relationship to the distal radius at the watershed line. Five VLP designs were fixed to each specimen based on their anatomic fit, and slid distally until the distal edge of the plate reached the watershed line. The position of each fixed plate was evaluated by fluoroscopy. We used a 3-dimensional laser scanner to create computer models. The total surface area of plate prominence volar to the watershed line and the prominent area beneath the FPL were measured in the axial plane using computer software.

RESULTS

At the watershed line, the FPL was located at 54% of the maximal width of the radius, as measured from its volar-ulnar corner. There were no significant differences in the location of plate fixation on lateral view radiographs according to the classification of Soong et al. The mean total surface area of plate prominence was 36 mm(2). The mean prominent area beneath the FPL was 10 mm2. Significant differences in plate prominence were noted for various designs.

CONCLUSIONS

Despite optimal plate placement, various VLP designs were observed to have prominent profiles volar to the watershed line, to varying extents.

CLINICAL RELEVANCE: The results raise concerns regarding interference between all of the analyzed VLP designs and the FPL. This study may help guide both implant design considerations and assist the surgeon in better understanding implant morphology as it relates to iatrogenic flexor tendon injury.

Для цитирования: Limthongthang Roongsak, Bachoura Abdo, Jacoby Sidney M, Osterman A Lee. Distal radius volar locking plate design and associated vulnerability of the flexor pollicis longus.. The Journal of hand surgery. 2014. DOI: 10.1016/j.jhsa.2014.01.038.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.