DOI 10.1111/j.1399-6576.1988.tb02751.x
Elimination of waste anaesthetic gases from operating theatres.
Аннотация
At present, waste anaesthetic gases in the operating theatre are eliminated by room ventilation and by additional measures such as scavenging by a double-mask system and a local exhaust system. A simple technique is proposed for measuring the capture efficiency (E) of different scavenging systems. E is the ratio of the waste anaesthetic gases captured by a system to the total quantity of waste anaesthetic gases produced. In a field study on existing unmodified system, E from a phantom pollutant source was measured using a portable measuring unit designed for displaying test results on location. The pollutant source was a leaking double-mask system. When scavenging from the double-mask system, E was 98% at an exhausted airflow of 28-33 m3/h. Using a local exhaust system at an airflow of 30 m3/h, E ranged from 20 to 96%, depending on the distance between the mask and the nozzle of the exhaust system.