A portable, 8-channel transcutaneous stimulator for paraplegic muscle training and mobility--a technical note.

Simcox Scott, Davis Glen, Barriskill Andrew, Middleton James, Bruinsma Ian, Duncan Michael, Smith Richard

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод в очереди

Аннотация

This paper introduces an 8-channel transcutaneous neuromuscular stimulator, called ExoStim, which was designed and developed to provide stimulation to the lower-limb muscles of spinal cord injured individuals. The intended purposes of the ExoStim were to act as a skin-surface precursor to an implantable neuromuscular stimulator for the specific tasks of increasing paralyzed leg strength and endurance, enabling the performance of basic lower-limb functional tasks, and familiarizing patients with functional electrical stimulation training. The initial design specifications included portability (approximately 500 g), battery-powered output, constant current control (0-300 mA), 8 channels of biphasic stimulation (charge-balanced, constant current), and microprocessor control of all stimulation parameters. Various tests, including output power characteristics, environmental, mechanical, and battery life, were performed on three prototype units to validate our design specifications. Having successfully passed all tests, the ExoStim is now ready to be deployed to clinical trial sites for further evaluation with spinal cord injured subjects.

Для цитирования: Simcox Scott, Davis Glen, Barriskill Andrew, Middleton James, Bruinsma Ian, Duncan Michael, Smith Richard. A portable, 8-channel transcutaneous stimulator for paraplegic muscle training and mobility--a technical note.. Journal of rehabilitation research and development. 2004. DOI: 10.1682/jrrd.2004.01.0041.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.