Evaluating the Effect of Flow and Interface Type on Pressures Delivered With Bubble CPAP in a Simulated Model.

Bailes Stephanie A, Firestone Kimberly S, Dunn Diane K, McNinch Neil L, Brown Miraides F, Volsko Teresa A

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Bubble CPAP, used for spontaneously breathing infants to avoid intubation or postextubation support, can be delivered with different interface types. This study compared the effect that interfaces had on CPAP delivery. We hypothesized that there would be no difference between set and measured levels between interface types.

METHODS

A validated preterm infant nasal airway model was attached to the ASL 5000 breathing simulator. The simulator was programmed to deliver active breathing of a surfactant-deficient premature infant with breathing frequency at 70 breaths/min inspiratory time of 0.30 s, resistance of 150 cm H2O/L/s, compliance of 0.5 mL/cm H2O, tidal volume of 5 mL, and esophageal pressure of -10 cm H2O. Nasal CPAP prongs, size 4030, newborn and infant RAM cannulas were connected to a nasal airway model and a bubble CPAP system. CPAP levels were set at 4, 5, 6, 7, 8, and 9 cm H2O with flows of 6, 8, and 10 L/min each. Measurements were recorded after 1 min of stabilization. The analysis was performed using SAS 9.4. The Kolmogorov-Smirnov test assessed normality of the data. The Friedman test was used to compare non-normally distributed repeated measures. The Wilcoxon signed-rank test was used to conduct post hoc analysis. All tests were 2-sided, and P values of <.05 were considered as indicating significant differences unless otherwise indicated.

RESULTS

At lower set CPAP levels, 4-6 cm H2O, measured CPAP dropped precipitously with the nasal prongs with the highest flow setting. At higher CPAP levels, 7-9 cm H2O measured CPAP concomitantly increased as the flow setting increased. Statistically significant differences in set and measured CPAP occurred for all devices across all CPAP levels, with the measured CPAP less than set for all conditions, P < .001.

CONCLUSIONS

Set flow had a profound effect on measured CPAP. The concomitant drop in measured pressure with high and low flows could be attributed to increased resistance to spontaneous breathing or insufficient flow to meet inspiratory demand. Clinicians should be aware of the effect that the interface and flow have on CPAP delivery.

Для цитирования: Bailes Stephanie A, Firestone Kimberly S, Dunn Diane K, McNinch Neil L, Brown Miraides F, Volsko Teresa A. Evaluating the Effect of Flow and Interface Type on Pressures Delivered With Bubble CPAP in a Simulated Model.. Respiratory care. 2015. DOI: 10.4187/respcare.04251.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.