DOI 10.1109/embc.2015.7320169

Медицинская библиотека · Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2015

An ECG oximetry system for identifying obstructive and central apnoea events.

de Chazal Philip, Sadr Nadi, Jayawardhana Madhuka

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

An automatic algorithm for processing simultaneously acquired electrocardiogram (ECG) and oximetry signals that identifies epochs of pure central apnoea, epochs containing obstructive apnoea and epochs of normal breathing is presented. The algorithm uses time and spectral features from the ECG derived heart-rate and respiration information, as well as features capturing desaturations from the oximeter sensor. Evaluation of performance of the system was achieved by using leave-one-record-out cross validation on the St. Vincent's University Hospital / University College Dublin Sleep Apnea Database from the Physionet collections of recorded physiologic signals. When classifying the three epoch types, our system achieved a specificity of 80%, a sensitivity to central apnoea of 44% and sensitivity to obstructive apnoea of 35%. A sensitivity of 81% was achieved when the central and obstructive epochs were combined into one class.

Для цитирования: de Chazal Philip, Sadr Nadi, Jayawardhana Madhuka. An ECG oximetry system for identifying obstructive and central apnoea events.. Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference. 2015. DOI: 10.1109/embc.2015.7320169.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.