DOI 10.1097/00126097-200512000-00009

Медицинская библиотека · Blood pressure monitoring · 2005

Validation of the TM-2564G device for in-hospital blood pressure measurement.

Kobalava Zhanna D, Kotovskaya Yulia V, Rodionov Eugeniy V, Moiseev Valentin S

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To perform clinical validation of the TM-2564G device for in-hospital blood pressure measurement according to the British Hypertension Society protocol.

METHODS

The tests were carried out on 85 patients aged 19-82 years (41 male patients, mean age 50+/-19 years) with systolic blood pressure 75-219 mmHg and diastolic blood pressure 43-117 mmHg (mean 142+/-33/85+/-20 mmHg). For each patient, three readings of the TM-2564G device were compared with simultaneous auscultatory measurements by two trained independent observers. The observers used a mercury-calibrated sphygmomanometer and dual stethoscope. The results were graded according to the British Hypertension Society protocol 1993.

RESULTS

The average difference was -1.85 mmHg for systolic blood pressure and -1.20 mmHg for diastolic blood pressure. The standard deviation of the difference between TM-2564G and the observers' results was 4.85 and 4.16 mmHg for systolic and diastolic blood pressures, respectively. Systolic blood pressure percentage of readings agreeing to within 5 mmHg was 75.7%, 10 mmHg 95.36% and 15 mmHg 99.6%, and for DBP 82.7, 97.6 and 99.8%, respectively.

CONCLUSIONS

The TM-2564G device achieved A/A British Hypertension Society grade and can be recommended for in-hospital measurement of blood pressure.

Для цитирования: Kobalava Zhanna D, Kotovskaya Yulia V, Rodionov Eugeniy V, Moiseev Valentin S. Validation of the TM-2564G device for in-hospital blood pressure measurement.. Blood pressure monitoring. 2005. DOI: 10.1097/00126097-200512000-00009.