A prospective randomized multicenter trial comparing clinical outcomes of patients treated surgically with a static or dynamic implant for acute ankle syndesmosis rupture.

Laflamme Mélissa, Belzile Etienne L, Bédard Luc, van den Bekerom Michel P J, Glazebrook Mark, Pelet Stéphane

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

To compare the clinical and radiographic outcome after stabilization of an acute syndesmosis rupture with either a static implant (a 3.5-mm metallic screw through 4 cortices) or a dynamic device (TightRope; Arthrex).

DESIGN

Multicenter randomized double-blind controlled trial.

SETTINGS: Study realized in 5 trauma centers (2 level 1 and 3 level 2) in 2 countries.

PATIENTS/

PARTICIPANTS

Seventy subjects admitted for an acute ankle syndesmosis rupture entered the study and were randomized into 2 groups (dynamic fixation = 34 and static fixation = 36). The 2 groups were similar regarding demographic, social, and surgical data. Sixty-five patients (dynamic = 33 and static = 32) completed the study and were available for analysis.

INTERVENTION: Syndesmosis fixation in the static group was realized with a 4 cortices 3.5-mm cortical screw (Synthes) and in the dynamic group with 1 TightRope (Arthrex). Standardized rehabilitation process for the 2 groups: no weight bearing in a cast for 6 weeks and then rehabilitation without protection.

MAIN OUTCOME MEASUREMENT: Olerud-Molander score.

RESULTS

Subjects with dynamic fixation achieved better clinical performances as described with the Olerud-Molander scores at 3 (68.8 vs. 60.2, P = 0.067), 6 (84.2 vs. 76.8, P = 0.082), and 12 months (93.3 vs. 87.6, P = 0.046). We also observed higher American Orthopaedic Foot and Ankle Society scores at 3 months (78.6 vs. 70.6, P = 0.016), but these were not significant at 6 (87.1 vs. 83.8, P = 0.26) or 12 months (93.1 vs. 89.9, P = 0.26). Implant failure was higher in the screw group (36.1% vs. 0%, P < 0.05). Loss of reduction was observed in 4 cases in the static screw group (11.1% vs. 0%, P = 0.06).

CONCLUSIONS

Dynamic fixation of acute ankle syndesmosis rupture with a dynamic device seems to result in better clinical and radiographic outcomes. The implant offers adequate syndesmotic stabilization without failure or loss of reduction, and the reoperation rate is significantly lower than with conventional screw fixation.

LEVEL OF EVIDENCE: Therapeutic Level II. See Instructions for Authors for a complete description of levels of evidence.

Для цитирования: Laflamme Mélissa, Belzile Etienne L, Bédard Luc, van den Bekerom Michel P J, Glazebrook Mark, Pelet Stéphane. A prospective randomized multicenter trial comparing clinical outcomes of patients treated surgically with a static or dynamic implant for acute ankle syndesmosis rupture.. Journal of orthopaedic trauma. 2015. DOI: 10.1097/bot.0000000000000245.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.