OBJECTIVE

To evaluate the impact of inflammation on anticoagulation monitoring for patients supported with extracorporeal membrane oxygenation (ECMO).

DESIGN

Prospective single-center cohort study.

SETTING

University-affiliated tertiary care academic medical center.

PARTICIPANTS

Adult venovenous and venoarterial ECMO patients anticoagulated with heparin/ MEASUREMENTS AND MAIN

RESULTS

C-Reactive protein (CRP) was used as a surrogate for overall inflammation. The relationship between CRP and the partial thromboplastin time (PTT, seconds) was evaluated using a CRP-insensitive PTT assay (PTT-CRP) in addition to measurement using a routine PTT assay. Data from 30 patients anticoagulated with heparin over 371 ECMO days was included. CRP levels (mg/dL) were significantly elevated (median, 17.2; interquartile range [IQR], 9.2-26.1) and 93% of patients had a CRP of ≥5. The median PTT (median 58.9; IQR, 46.9-73.3) was prolonged by 11.3 seconds compared with simultaneously measured PTT-CRP (median, 47.6; IQR, 40.1-55.5; p < 0.001). The difference between PTT and PTT-CRP generally increased with CRP elevation from 2.7 for a CRP of <5.0 to 13.0 for a CRP between 5 and 10, 17.7 for a CRP between 10 and 15, and 15.1 for a CRP of >15 (p < 0.001). In a subgroup of patients, heparin was transitioned to argatroban, and a similar effect was observed (median PTT, 62.1 seconds [IQR, 53.0-78.5 seconds] vs median PTT-CRP, 47.6 seconds [IQR, 41.3-57.7 seconds]; p < 0.001).

CONCLUSIONS

Elevations in CRP are common during ECMO and can falsely prolong PTT measured by commonly used assays. The discrepancy due to CRP-interference is important clinically given narrow PTT targets and may contribute to hematological complications.

Для цитирования: Madhok Jai, O'Donnell Christian, Jin Jing, Owyang Clark G, Weimer Jonathan M, Pashun Raymond A, Shudo Yasuhiro, McNulty John, Chadwick Blaine, Ruoss Stephen J, Rao Vidya K, Zehnder James L, Hsu Joe L. Impact of C-reactive Protein on Anticoagulation Monitoring in Extracorporeal Membrane Oxygenation.. Journal of cardiothoracic and vascular anesthesia. 2024. DOI: 10.1053/j.jvca.2024.04.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.