The CLOSED protocol to assess optic nerve sheath diameter using color-Doppler: a comparison study in a cohort of idiopathic normal pressure hydrocephalus patients.
Аннотация
BACKGROUND
Sonographic assessment of the optic nerve sheath diameter represents a promising non-invasive technique for estimation of the intracranial pressure. A wide inter-observer variability, along with a lack of a standardized protocol for the optic nerve sheath diameter measurements, could lead to over- or under-estimation. The present study was aimed at evaluating feasibility of color-Doppler for better delineating optic nerve sheath borders, comparing it to B-mode imaging, using the magnetic resonance measurements as a comparison.
METHODS
Optic nerve sheath diameters were evaluated using magnetic resonance by an expert radiologist in a cohort of patients with suspected idiopathic normal pressure hydrocephalus. Magnetic resonance findings were evaluated twice. In the first half of this cohort, optic nerve sheath diameters were measured using B-mode only, in the second half applying color-Doppler. Measurements obtained using these two techniques were compared to magnetic resonance imaging measurements. The Bland-Altman analysis and concordance correlation coefficient were computed to quantify the strength of agreement between the two magnetic resonance assessments. Box plots and average (± SD) were used to compare assessments by sonographic and magnetic resonance methods.
RESULTS
Fifty patients were included. MRI assessment showed a moderate concordance correlation coefficient. Optic nerve sheath diameters measured applying color-Doppler were lower (p < 0.001) and less scattered compared to B-mode assessment, which approached more to magnetic resonance measurements.
CONCLUSIONS
In this cohort of patients, magnetic resonance showed high intra-rater variability in optic nerve sheath diameter assessments. Optic nerve sheath diameter assessments using color-Doppler yielded lower and less scattered diameters compared to B-mode only.