Аннотация

OBJECTIVES

To measure the extent of dilution of helium-oxygen (heliox) by room air when given via high concentration reservoir mask to spontaneously breathing subjects. Substantial dilution of heliox by room air under these circumstances might alter its physical properties sufficiently to negate any potential clinical benefit in obstructive respiratory failure.

DESIGN

Healthy volunteers breathing different concentrations of helium in oxygen via two different masks in a randomised crossover design.

SETTING

Operating theatre in a university hospital.

PATIENTS AND

PARTICIPANTS

Six healthy volunteers.

INTERVENTIONS: The concentrations of helium, nitrogen and oxygen were measured in the trachea of each volunteer using a mass spectrometer during normal breathing, hyperventilation and hypoventilation.

MEASUREMENTS AND

RESULTS

During normal breathing of Heliox21 (79% helium) via a standard non-rebreathe reservoir mask, within subject median percentage tracheal helium was 37.2% (range 29.3-52.2%) and nitrogen was 41.7% (27.4-49.4%). Air entrainment was affected by changes in breathing pattern: tracheal nitrogen concentration was greater during hyperventilation (55.4%; range 49.4-63.5%) and less during hypoventilation (33.1%; range 24.6-39.6%, p=0.043). Tracheal nitrogen could be almost completely abolished by administering heliox via a tightly fitting cushioned facemask, even during hyperventilation (2.2%; range 0.6-6.1%, p=0.028).

CONCLUSIONS

Heliox administration via a standard high-concentration reservoir mask leads to significant dilution by room air. For the full potential benefits of heliox to be realised in spontaneously breathing patients, it should be administered via a system that achieves a gas tight seal, with no leaks between the delivery device and the surroundings.

Для цитирования: Standley Thomas D A, Smith Helen L, Brennan Liam J, Wilkins Ingrid A, Bradley Peter G, Barrera Groba Casiano, Davey Andrew J, Menon David K, Wheeler Daniel W. Room air dilution of heliox given by facemask.. Intensive care medicine. 2008. DOI: 10.1007/s00134-008-1064-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.