Procedural and cardiovascular outcomes of geriatric vs non-geriatric patients undergoing permanent pacemaker implantation - a nationwide cohort analysis.
Аннотация
BACKGROUND
Permanent pacemaker implantation is increasing exponentially to treat atrio-ventricular block and symptomatic bradyarrhythmia. Despite being a minor surgery, immediate complications such as pocket infection, pocket hematoma, pneumothorax, hemopericardium, and lead displacement do occur.
METHODS
The Nationwide Inpatient Sample was queried from 2016 to 2018 to identify patients with pacemakers using ICD-10 procedure code. The Chi-square test was used for statistical analysis.
RESULTS
The sample size consisted of 443,460 patients with a pacemaker, 26% were <70 years (male 57%, mean age of (60.6±9.7) yr, Caucasian 70%) and 74% were ≥70 years (male 50%, mean age of (81.4±5.9) yr, Caucasian 79%). Upon comparison of rates in the young vs elderly: mortality (1.6% vs 1.5%; P<0.01), obesity (26% vs 13%; P<0.001), coronary artery disease (40% vs 49%; P<0.001), HTN (74% vs 87%; P<0.01), anemia (4% vs 5%; P<0.01), atrial fibrillation (34% vs 49%; P<0.01), peripheral artery disease (1.7% vs 3%; P<0.01), CHF (31% vs 39%; P<0.001), diabetes (31% vs 27.4%; P<0.01), vascular complications (1.1% vs 1.2%; P<0.01), pocket hematoma (0.5% vs 0.8%; P<0.01), AKI (16% vs 21%; P<0.01), hemopericardium (0.1% vs 0.1%; P = 0.1), hemothorax (0.3% vs 0.2%; P<0.01), cardiac tamponade (0.4% vs 0.5%; P<0.01), pericardiocentesis (0.4% vs 0.4%; P<0.01), cardiogenic shock (4% vs 2.3%; P<0.01), respiratory complications (1.9% vs 0.9%; P<0.01), mechanical ventilation (5.1% vs 2.9%; P<0.01); post-op bleed (0.5% vs 0.3%; P<0.01), need for transfusion (4.8% vs 3.8%; P<0.01), severe sepsis (0.6% vs 0.5%; P<0.01 ), septic shock (2% vs 1%; P<0.01), bacteraemia (0.8% vs 0.4%; P<0.01), lead dislodgement (1.4% vs 1.1%; P<0.01).
CONCLUSIONS
Our study revealed that the overall complication rates were lower in the elderly despite higher co-morbidities. This aligns with previous studies which showed lower rates in the elderly. Hence providers should not hesitate to provide guideline driven pacemaker placement in the elderly especially in patients with good life expectancy.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.