Contributors to and impact of residual shunting after device closure of atrial septal defects.
Аннотация
BACKGROUND
The prevalence of residual shunt in patients after device closure of atrial septal defect and its impact on long-term outcome has not been previously defined.
METHODS
From a prospective, single-institution registry of 408 patients, we selected individuals with agitated saline studies performed 1 year after closure. Baseline echocardiographic, invasive hemodynamic, and comorbidity data were compared to identify contributors to residual shunt. Survival was determined by review of the medical records and the Social Security Death Index. Survival analysis according to shunt included construction of Kaplan-Meier curves and Cox proportional hazards modeling.
RESULTS
Among 213 analyzed patients, 27% were men and age at repair was 47 ± 17 years. Thirty patients (14%) had residual shunt at 1 year. Residual shunt was more common with Helex (22%) and CardioSEAL/STARFlex (40%) occluder devices than Amplatzer devices (9%; P = .005). Residual shunts were more common in whites (79% vs 46%, P = .004). At 7.3 ± 3.3 years of follow-up, 13 (6%) of patients had died, including 8 (5%) with Amplatzer, 5 (25%) with CardioSEAL/STARFlex, and 0 with Helex devices. Patients with residual shunting had a higher hazard of death (20% vs 4%, P = .001; hazard ratio 4.95 [1.59-14.90]). In an exploratory multivariable analysis, residual shunting, age, hypertension, coronary artery disease, and diastolic dysfunction were associated with death.
CONCLUSIONS
Residual shunt after atrial septal defect device closure is common and adversely impacts long-term survival.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.