DOI 10.1097/01.ijg.0000159120.03747.48

Медицинская библиотека · Journal of glaucoma · 2005

Measurement of intraocular pressure using the NT-4000: a new non-contact tonometer equipped with pulse synchronous measurement function.

Yaoeda Kiyoshi, Shirakashi Motohiro, Fukushima Atsushi, Funaki Shigeo, Funaki Haruko, Ofuchi Nobutaka, Nakatsue Tomoko, Abe Haruki

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: NT-4000 (Nidek Co. Ltd., Gamagori, Japan) is a new non-contact tonometer (NCT) equipped with pulse synchronous measurement function that can measure intraocular pressure (IOP) synchronized with the ocular pulse. The purpose of this study was to evaluate the usefulness of NT-4000 in normal subjects and in patients with glaucoma and ocular hypertension.

METHODS

This study included 175 eyes of 175 subjects. Firstly, the IOP was measured using NT-4000 without the pulse synchronous measurement function (NTn). Secondly, the IOP at peak, middle, and trough phases of the pulse signal were measured using NT-4000 with the pulse synchronous measurement function (NTp, NTm, NTt, respectively). Additionally, the IOP was measured with Goldmann applanation tonometer (GT). The coefficient of variation (CV) of three readings in the NCT measurements was used to evaluate the intra-session reproducibility. Statistical comparisons were performed using Wilcoxon signed rank test and one-way analysis of variance with Scheffe's test. Linear regression analysis was used to calculate correlation coefficients. P values less than 0.05 were accepted as statistically significant.

RESULTS

The CV of NTn, NTp, NTm, and NTt were 6.4%, 5.5%, 4.9%, and 5.2%, respectively. The CV of NTp, NTm, and NTt were significantly smaller than that of NTn (P = 0.007, P < 0.001, P < 0.001, respectively). NTp was significantly higher than NTt (P = 0.038). GT was significantly correlated with NTn, NTp, NTm, and NTt (r = 0.898, P < 0.001; r = 0.912, P < 0.001; r = 0.908, P < 0.001; r = 0.900, P < 0.001, respectively).

CONCLUSION

NT-4000 can detect the fluctuation of IOP associated with the ocular pulse.

Для цитирования: Yaoeda Kiyoshi, Shirakashi Motohiro, Fukushima Atsushi, Funaki Shigeo, Funaki Haruko, Ofuchi Nobutaka, Nakatsue Tomoko, Abe Haruki. Measurement of intraocular pressure using the NT-4000: a new non-contact tonometer equipped with pulse synchronous measurement function.. Journal of glaucoma. 2005. DOI: 10.1097/01.ijg.0000159120.03747.48.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.