MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2147/cia.s130882

Медицинская библиотека · Clinical interventions in aging · 2017

The safety issues and hardware-related complications of deep brain stimulation therapy: a single-center retrospective analysis of 478 patients with Parkinson's disease.

Zhang Jing, Wang Tao, Zhang Chen-Cheng, Zeljic Kristina, Zhan Shikun, Sun Bo-Min, Li Dian-You

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Deep brain stimulation (DBS) is a well-established therapy for the treatment of advanced Parkinson's disease (PD) in patients experiencing motor fluctuations and medication-refractory tremor. Despite the relative tolerability and safety of this procedure, associated complications and unnatural deaths are still unavoidable.

METHODS

In this study, hardware-related complications and the causes of unnatural death were retrospectively analyzed in 478 patients with PD who were treated with DBS.

RESULTS

The results showed a 3-year survival rate of 98.6% and a 5-year survival rate of 96.4% for patients with PD who underwent DBS treatment at the study center. Pneumonia was the cause of death with the highest frequency. Prophylactic antibiotics and steroids or antihistamine drugs were adopted to reduce the risk of infection. Twenty-two patients (4.6%) experienced hardware-related complications.

CONCLUSION

Deaths of PD patients who receive DBS are typically unrelated to the disease itself or complications associated with the surgery. Pneumonia, malignant tumors, asphyxia, and multiple-organ failure are the common causes of death. Swallowing-related problems may be the most important clinical symptom in late-stage PD, as they cannot be stabilized or improved by DBS alone, and are potentially lethal. Although prophylactic antibiotics and steroids or antihistamine drugs may reduce the risk of infection, it is imperative to identify high-risk patients for whom a therapeutic approach not requiring an implantable device is more suitable, for example, pallidotomy and potentially transcranial ultrasound.

Для цитирования: Zhang Jing, Wang Tao, Zhang Chen-Cheng, Zeljic Kristina, Zhan Shikun, Sun Bo-Min, Li Dian-You. The safety issues and hardware-related complications of deep brain stimulation therapy: a single-center retrospective analysis of 478 patients with Parkinson's disease.. Clinical interventions in aging. 2017. DOI: 10.2147/cia.s130882.