Adverse Events and Device Failures Associated with Pancreatic Stents: A Comprehensive Analysis Using the FDA's MAUDE Database.

Jaber Fouad, Alsakarneh Saqr, Alsharaeh Tala, Salahat Ahmed-Jordan, Jaber Mohammad, Abboud Yazan, Aldiabat Mohammad, Ahmed Khalid, Abdallah Mohamed, Abdel-Jalil Ala, Bilal Mohammad, Barakat Monique, Adler Douglas G

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Pancreatic duct stents (PDS) are widely used for the prevention of post-endoscopic retrograde cholangiopancreatography (ERCP) pancreatitis. However, there is a paucity of data regarding the adverse events associated with PDS placement. This study aims to investigate the reported adverse events and device failures related to PDS, utilizing the Manufacturer and User Facility Device Experience (MAUDE) database maintained by the U.S. Food and Drug Administration (FDA).

METHODS

Post-marketing surveillance data from January 2013 to December 8, 2023, were extracted from the FDA's MAUDE database to analyze the reports pertaining to the use of commonly used PDS. The primary outcomes of interest were device issues and patient-related adverse events. Statistical analysis was performed using Microsoft Excel 2010, with the calculation of pooled numbers and percentages for each device and patient adverse event.

RESULTS

A total of 579 device issues and 194 patient-related adverse events were identified. Device issues were primarily attributed to stent deformation (n = 72; 12.4%), followed by migration of the device into the pancreatic duct or expulsion out of the duct (n = 60; 10.4%), and stent fracture/breakage (n = 55; 9.4%). Among the patient-reported adverse events, inflammation was the most common (n = 26; 13.4%), followed by reports of stents becoming embedded in tissue (n = 21; 10.8%) and stent occlusion/obstruction (n = 16; 8.2%). The most prevalent device failures associated with Advanix stents were material deformation, with perforation (n = 3, 30%) being the most frequently reported adverse event. Concerning Geenen stents, migration or expulsion of the device (n = 34, 16.9%) constituted the most common device-related adverse events, while inflammation (n = 20, 16.7%) was the most frequently reported patient-related issue. For Zimmon stents, migration or expulsion of the device (n = 22, 8.8%) were the most frequently reported device-related problems, whereas perforation (n = 7, 10.9%) and bleeding (n = 7, 10.9%) were the most frequent patient-related adverse events.

CONCLUSION

Our findings highlight important device and patient adverse events that endoscopists and referring providers should be aware of before considering pancreatic stent placement.

Для цитирования: Jaber Fouad, Alsakarneh Saqr, Alsharaeh Tala, Salahat Ahmed-Jordan, Jaber Mohammad, Abboud Yazan, Aldiabat Mohammad, Ahmed Khalid, Abdallah Mohamed, Abdel-Jalil Ala, Bilal Mohammad, Barakat Monique, Adler Douglas G. Adverse Events and Device Failures Associated with Pancreatic Stents: A Comprehensive Analysis Using the FDA's MAUDE Database.. Digestive diseases and sciences. 2024. DOI: 10.1007/s10620-024-08396-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.