DOI 10.1016/j.ijcard.2023.131165

Медицинская библиотека · International journal of cardiology · 2023

Assay interference as a cause of false positive troponin T elevation in emergency department patients.

Lyons K S, Herity N, Lee G, Talbot C, McKeeman G

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Troponin assays are used in the diagnosis of myocardial injury and may show elevated results for a variety of reasons. However it is increasingly recognised that cardiac troponin elevation may in some cases be due to assay interference. This is of significant importance as a misdiagnosis of myocardial injury may lead to unnecessary and potentially harmful investigation and treatment for patients. We sought to confirm the accuracy of cardiac high sensitivity troponin T (chsTnT) elevation in an unselected group of patients presenting to the emergency department, by using a second confirmatory cardiac high sensitivity troponin I (chsTnI) assay.

METHODS

We identified patients presenting to two local emergency departments over a five-day period who had chsTnT levels measured as part of routine clinical care. All samples with elevated chsTnT levels (above the 99% centile URL) were retested for chsTnI in order to confirm true myocardial injury.

RESULTS

A total of 74 samples from 54 patients were analysed for chsTnT and chsTnI. 7 samples (9.5%) had chsTnI levels < 5 ng/L suggesting assay interference as the cause of chsTnT elevation.

CONCLUSIONS

Assay interference leading to false positive troponin elevation may be more common than many physicians appreciate and can potentially lead to harmful investigation and treatment for patients. In cases where the diagnosis of myocardial injury is uncertain, a second alternative troponin assay should be performed to confirm true myocardial injury.

Для цитирования: Lyons K S, Herity N, Lee G, Talbot C, McKeeman G. Assay interference as a cause of false positive troponin T elevation in emergency department patients.. International journal of cardiology. 2023. DOI: 10.1016/j.ijcard.2023.131165.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.