The reliability of portable fixed dynamometry during hip and knee strength assessments.

Kollock Roger O, Onate James A, Van Lunen Bonnie

Клиническое исследованиеСтатьяВалидационное исследованиеПеревод в очереди

Аннотация

CONTEXT: Insufficient lower extremity strength may be a risk factor for lower extremity injuries such as noncontact anterior cruciate ligament tears. Therefore, clinicians need reliable instruments to assess strength deficiencies.

OBJECTIVE

To assess the intrarater, interrater, intrasession, and intersession reliability of a portable fixed dynamometer in measuring the strength of the hip and knee musculature.

DESIGN

Crossover study.

SETTING

Sports medicine research laboratory.

PATIENTS OR OTHER

PARTICIPANTS

Three raters (A, B, C) participated in this 2-phase study. Raters A and B tested 11 healthy college graduate students (2 men, 9 women) in phase 1. Raters A and C tested 26 healthy college undergraduate students (7 men, 19 women) in phase 2.

MAIN OUTCOME MEASURE(S): The dependent variables for the study were hip adductor, hip abductor, hip flexor, hip extensor, hip internal rotator, hip external rotator, knee flexor, and knee extensor peak force.

RESULTS

The phase 1 intrasession intraclass correlation coefficients for sessions 1, 2, and 3 ranged from 0.88 to 0.99 (SEM = 0.08-3.02 N), 0.85 to 0.99 (SEM = 0.26-3.88 N), and 0.92 to 0.96 (SEM = 0.52-2.76 N), respectively. Intraclass correlation coefficients ranged from 0.57 to 0.95 (SEM = 1.72-13.15 N) for phase 1 intersession values, 0.70 to 0.94 (SEM = 1.42-9.20 N) for phase 2 intrarater reliability values, and 0.69 to 0.88 (SEM = 1.20-8.50 N) for phase 2 interrater values.

CONCLUSIONS

The portable fixed dynamometer showed good to high intrasession and intersession reliability values for hip and knee strength. Intrarater and interrater reliability were fair to high, except for hip internal rotation, which showed poor reliability.

Для цитирования: Kollock Roger O, Onate James A, Van Lunen Bonnie. The reliability of portable fixed dynamometry during hip and knee strength assessments.. Journal of athletic training. 2010. DOI: 10.4085/1062-6050-45.4.349.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.