Successful retrieval of a Figulla Occlutech septal occluder - embolized device stability and potential solutions.
Клинический случайСтатьяПеревод в очереди
Аннотация
Go to: We report a case of an atrial septal defect where a Figulla Occlutech device embolized into the right ventricle. As this device has no left atrial hub, we brought the bare device into the inferior vena cava and used a novel two-snare technique to slenderize the device into the sheath. This report highlights additional tips in transcatheter device retrieval.
Для цитирования: Sadiq Masood, Ur Rehman Asif, Mehmood Amjad. Successful retrieval of a Figulla Occlutech septal occluder - embolized device stability and potential solutions.. Annals of pediatric cardiology. 2016. DOI: 10.4103/0974-2069.180672.