Journal of clinical microbiology · 2002
Evaluation of a new combined antigen and antibody human immunodeficiency virus screening assay, VIDAS HIV DUO Ultra.
Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Weber Bernard, Berger Annemarie, Rabenau Holger, Doerr Hans Wilhelm
Аннотация
Combined antigen and antibody screening (fourth-generation) assays reduce the diagnostic window period between the time of human immunodeficiency virus (HIV) infection and laboratory diagnosis by 4 days, on average, in comparison to antibody-only (third generation) enzyme immunoassays (EIAs). The aim of the present study was to assess whether the new VIDAS HIV DUO Ultra (Biomérieux, Marcy-l'Etoile, France) showed an improved sensitivity and specificity in comparison to licensed fourth-generation assays. A total of 16 seroconversion panels, 15 cell culture supernatants infected with different HIV type 1 (HIV-1) subtypes, and 257 potentially cross-reactive serum samples were tested with VIDAS DUO HIV Ultra, Genscreen Plus HIV Ag-Ab, Enzygnost HIV Integral, Enzymun-Test HIV Combi, Genscreen HIV 1/2, version 2 (third-generation EIA), and Genetic Systems HIV-1 Ag EIA (p24 antigen assay). VIDAS HIV DUO Ultra showed a comparable sensitivity to the single p24 antigen assay in seroconversion panels and a dilution series of virus lysates. The diagnostic window was reduced with VIDAS HIV DUO Ultra by 3.82 days, on average, in comparison with the fourth-generation assay with the lowest sensitivity of the antigen detection module. HIV-1 infection was detected 5.88 days earlier than with third-generation EIA. The mean time delay between reverse transcription-PCR and VIDAS HIV DUO Ultra was only 2.31 days. The specificity of fourth-generation assays after retesting ranged between 98.1 and 100%. In conclusion, VIDAS HIV DUO Ultra can replace single-antigen screening for laboratory diagnosis and screening of HIV infection in blood donors. There was no evidence for a second diagnostic window due to impaired sensitivity of the antibody detection module of all the fourth-generation EIAs evaluated in the present study. The specificity after initial and/or repeated testing of VIDAS HIV DUO Ultra was equivalent to that of a third-generation assay.