Repeatability and reproducibility of maximum diameter measurements of prostate lesions on MRI with repositioning and variation of imaging sequences: A test-retest study.

Zhang Kevin Sun, Glemser Philip Alexander, Neelsen Christian Jan Oliver, Wennmann Markus, Rotkopf Lukas Thomas, Netzer Nils, Meinzer Clara, Hielscher Thomas, Weru Vivienn, Görtz Magdalena, Stenzinger Albrecht, Hohenfellner Markus, Schlemmer Heinz-Peter, Bonekamp David

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To assess variability of maximum diameter measurements of prostate lesions in MRI assessing patient repositioning, rater and sequence effects.

METHODS

Forty-two patients were included retrospectively, who received a clinical bi-/multiparametric prostate MRI examination and agreed to have the T2-weighted (T2WI) and diffusion weighted-imaging (DWI) sequences scanned twice. Maximum diameter measurements of prostate lesions mentioned in the clinical radiologist reports were performed by four readers in multiple reading sessions for determination of inter-sequence (between two DWI sequences), inter-scan (between clinical and additional scan), intra-rater and inter-rater variability. The primary calculated metrics were the repeatability and reproducibility coefficient (RC/RDC), including pooled RC/RDC.

RESULTS

Variability measured by RCs/RDCs was lowest for measurements obtained within the same reading session, with inter-scan RCs up to 5.6 mm/6.5 mm for T2WI/DWI, pooled RCs of 4.8 mm/5.8 mm, respectively, and inter-sequence RDCs of 5.4 mm-5.9 mm, pooled RDC 5.8 mm. Measurements performed in separate reading sessions demonstrated significantly higher variability for both settings in the majority of cases (RCs: up to 10.9 mm/11.7 mm/10.2 mm for T2WI/DWI/inter-sequence, p ≤ 0.002), pooled RCs/RDCs 9.2 mm-9.9 mm. Measurements necessarily generated in different reading sessions, i.e., intra-rater or inter-rater, demonstrated high variability (RCs/RDCs up to 11.4 mm/11.5 mm for T2WI/DWI).

CONCLUSIONS

Prostate lesion measurements demonstrate considerable variability. When measured in one reading session by one rater, lesion diameter differences below the pooled RCs of 4.8 mm, 95 %-CI [3.9, 5.6] for T2WI and 5.8 mm, 95 %-CI [4.7, 7.1] for DWI should not necessarily assumed to be true biological change, as these differences may result from measurement- or repositioning-based variability alone. Caution needs to be taken assessing size changes.

Для цитирования: Zhang Kevin Sun, Glemser Philip Alexander, Neelsen Christian Jan Oliver, Wennmann Markus, Rotkopf Lukas Thomas, Netzer Nils, Meinzer Clara, Hielscher Thomas, Weru Vivienn, Görtz Magdalena, Stenzinger Albrecht, Hohenfellner Markus, Schlemmer Heinz-Peter, Bonekamp David. Repeatability and reproducibility of maximum diameter measurements of prostate lesions on MRI with repositioning and variation of imaging sequences: A test-retest study.. Magnetic resonance imaging. 2025. DOI: 10.1016/j.mri.2025.110578.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.